Skip to content

Turbett Surgical Container, TS1500 recallZ-1501-2022 · Turbett Surgical, Inc. · initiated Apr 14, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Turbett Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices were subject to unapproved rework processes.

FDA root cause category: Process control

What the company is doing

The firm distributed recall notification to hospitals via email on 04/14/2022. The recall notification identifies the reason for recall of specific TS1500 units to be that the repair process employed could restrict the customer's ability to clean and visually check the device for cleanliness. This can be identified by the customer by evidence of bioburden or particulate buildup in the seams of the device. Inadequate or incomplete cleaning of the unit may compromise the sterility of the load. The firm asks customers to immediately discontinue use of the device and complete the provided Return Response Form to [email protected]. Turbett will contact customers to coordinate the replacement of the unit and correction of the recalled device. Customers with any questions can call 585-755-0133x106 from Monday through Friday, from 8:00 AM to 4:30 PM.

Recall details

Recall number
Z-1501-2022
Status
Terminated
Initiated by the company
Apr 14, 2022
Posted by the FDA
Aug 4, 2022
Recalling firm
Turbett Surgical, Inc.
Quantity
7 units
Product code
FRG
FDA submissions
K153025
Distribution
US Nationwide Nationwide: CO, FL, NC and PA.

Product description

Turbett Surgical Container, TS1500

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 00859464007113; Lot No./Serial No.: 100036/2017061015, 100036/2017061016, 100036/2017061017, 100036/2017061019, 100036/2017061023, 100036/2017061025, 100065/2017121001 and 100081/2018031004.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.