Medical device recalls: August 2022
The FDA posted 98 medical device recall records in August 2022, from 16 recall events by 15 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- GE Healthcare, LLC · 2 events, 2 records
- Mckesson Medical-Surgical Inc. Corporate Office · 1 event, 55 records
- Family Dollar Stores, Llc. · 1 event, 22 records
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · 1 event, 6 records, 6 Class I
- Siemens Medical Solutions USA, Inc · 1 event, 3 records
- Baxter Healthcare Corporation · 1 event, 1 record
- Boston Scientific Corporation · 1 event, 1 record
- Heartware, Inc. · 1 event, 1 record, 1 Class I
FDA root cause
- Environmental control · 55
- Storage · 22
- Device Design · 10
- Under Investigation by firm · 3
- Process control · 3
- Process design · 2
- Process change control · 1
- Nonconforming Material/Component · 1
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- INTERA 3000 Hepatic Artery Infusion Pump, 30 mL, Sterile, Rx Only. implantable infusion pump.Class IZ-1522-2022 · Intera Oncology, Inc. · posted Aug 22, 2022 · Terminated
Higher than expected flow rate.
- Z-1561-2022 · Synapse Biomedical Inc · posted Aug 22, 2022 · Terminated
Due to a defect identified on the EPG controller printed circuit board assembly, device performance may degrade over time. Depending on the failure mode, potential adverse events include pain, muscle fatigue and decreased breathing effectiveness,…
- Z-1512-2022 · Heartware, Inc. · posted Aug 18, 2022 · Open, Classified
Battery performance issues. Battery electrical faults render it unable to power the controller, unable to accept charge from the battery charger, and/or result in the battery to appear to remain charged while discharging.
- Cobalt Implantable Cardioverter Defibrillators: a. Cobalt XT VR ICD, Model Numbers: DVPA2D1, DVPA2D4Class IZ-1485-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Aug 9, 2022 · Open, Classified
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
- Z-1486-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Aug 9, 2022 · Open, Classified
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
- Cobalt Implantable Cardioverter Defibrillators: a. Cobalt VR ICD, Model Numbers: DVPB3D1, DVPB3D4Class IZ-1487-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Aug 9, 2022 · Open, Classified
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
- Z-1488-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Aug 9, 2022 · Open, Classified
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
- Crome Implantable Cardioverter Defibrillators: a. Crome VR ICD, Model Numbers: DVPC3D1, DVPC3D4Class IZ-1489-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Aug 9, 2022 · Open, Classified
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
- Z-1490-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · posted Aug 9, 2022 · Open, Classified
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Class II recalls (89)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1648-2022 · Siemens Healthcare Diagnostics, Inc. · posted Aug 29, 2022 · Open, Classified
Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.
- Z-1583-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1584-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1585-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1586-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1587-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1588-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Drugs of Abuse Tests:Class IIZ-1589-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1590-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1591-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1592-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1593-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1594-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1595-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1596-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1597-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Strep test:Class IIZ-1598-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1599-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- MERIT MEDICAL SYSTEMS Temno Soft Tissue Biopsy Device Temno 14 Gauge 6 cm Length Beveled Sharp TipClass IIZ-1600-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1601-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1602-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1603-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1604-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1605-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1606-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1607-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1608-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- MEDLINE INDUSTRIES Medline Catheter Insertion Tray Medline Foley Without Catheter Without BalloonClass IIZ-1609-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1610-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1611-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1612-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1613-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1614-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1615-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1616-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1617-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1618-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1619-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1620-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1621-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1622-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1623-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1624-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1625-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1626-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1627-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
XXX
- Z-1628-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- BECKMAN COULTER, INC.Class IIZ-1629-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1630-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1631-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1632-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1633-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1634-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1635-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1636-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1637-2022 · Mckesson Medical-Surgical Inc. Corporate Office · posted Aug 26, 2022 · Open, Classified
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-1645-2022 · Siemens Medical Solutions USA, Inc · posted Aug 26, 2022 · Open, Classified
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any…
- Artis pheno- Model No. 10849000Class IIZ-1646-2022 · Siemens Medical Solutions USA, Inc · posted Aug 26, 2022 · Open, Classified
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any…
- Z-1647-2022 · Siemens Medical Solutions USA, Inc · posted Aug 26, 2022 · Open, Classified
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any…
- Z-1574-2022 · Medtronic Neuromodulation · posted Aug 25, 2022 · Open, Classified
Potential communication issue in which the Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 may be unable to connect with a Vanta Implantable Neurostimulator (INS) Model 977006 when a patient is implanted with more than one neurostimulator.
- Z-1582-2022 · Boston Scientific Corporation · posted Aug 25, 2022 · Open, Classified
The impacted products were distributed with an MPXF tip curve shape instead of the 40XF tip curve shape for the Guider/40XF/8FR/90CM guide catheter.
- Z-1566-2022 · GE Healthcare, LLC · posted Aug 24, 2022 · Open, Classified
Inaccurate Distance and Area measurements with use of Global Stack viewport.
- Z-1570-2022 · Baxter Healthcare Corporation · posted Aug 24, 2022 · Open, Classified
There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets.
- Z-1565-2022 · Nobel Biocare · posted Aug 23, 2022 · Open, Classified
Due to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.
- Z-1526-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1527-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1528-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1529-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1530-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1531-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1532-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1533-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1534-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1535-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1536-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1537-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1538-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1539-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1540-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1541-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1542-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1543-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1544-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1545-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1546-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1547-2022 · Family Dollar Stores, Llc. · posted Aug 17, 2022 · Open, Classified
Affected product was not appropriately stored in temperature-controlled areas between April 30, 2022 - June 10, 2022.
- Z-1511-2022 · GE Healthcare, LLC · posted Aug 10, 2022 · Open, Classified
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
- Z-1502-2022 · Neotract Inc · posted Aug 5, 2022 · Open, Classified
There is the potential that during implant deployment, the device may not properly deliver a implant.
- Z-1501-2022 · Turbett Surgical, Inc. · posted Aug 4, 2022 · Terminated
Devices were subject to unapproved rework processes.