VIZIGO recallZ-1677-2022 · Biosense Webster, Inc. · initiated Jun 29, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised sterile barrier could be a source for infection.

FDA root cause category: Material/Component Contamination

What the company is doing

On 6/29/22, re call notices started being emailed and mailed to customers who were asked to do the following: 1) The notice should be forwarded to anyone in the facility that needs to be aware. 2) Identify affected devices, unbox devices, and destroy the devices. 3) Complete and return the Business Reply Form and send back the label on empty boxes to receive credit. 4) If any of the affected products have been forwarded to another facility, contact that facility, and provide a copy of this letter to the relevant personnel. 5) Maintain a copy of this communication where the product identified in this letter is located until all products have been destroyed. If you have additional questions about this letter, please email the recalling firm at [email protected]

Recall details

Recall number
Z-1677-2022
Status
Open, Classified
Initiated by the company
Jun 29, 2022
Recalling firm
Biosense Webster, Inc.
Quantity
98
Product code
DYB
FDA submissions
K170997
Distribution
US Nationwide distribution in the states of AL, GA, AZ, TX, MA, VA, NY, ME, FL, IA, NM, NH, TN, WI, OH, SC, CA, IN.

Product description

CARTO VIZIGO 8.5F Bi-Directional Guiding Sheath, REF: D138501. Used to introduce various cardiovascular catheters into the heart.

Code information (models, lots, serial numbers)

UDI-DI: 10846835016253, Lot Number: 00002001

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K170997 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page