Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In… recallZ-1648-2022 · Siemens Healthcare Diagnostics, Inc. · initiated Jul 18, 2022
Class II Open, Classified All recalls posted in August 2022
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued US Customers Urgent Medical Device Correction (VC-22-02.A.US) and OUS Customers Urgent Field Safety Notice (VC-22-02.A.OUS) dated July 2022 to inform them to use lot specific correlation factors with all in-date lots of Dimension Vista CA15-3. Letter states reason for recall, health risk and action to take: Calibrate/Recalibrate Dimension Vista LOCI CA 15-3 lots listed in Table 1 using any in date lot of Dimension Vista LOCI 7 Calibrator (SMN 10488446/ REF KC605). Enter lot specific correlation factors C0 and C1 in the Method Configuration screen per Flex reagent Lot listed below in Table 2. After applying correlation factors to the specified lot, product claims listed in the Instructions For Use (IFU) are met For instructions on entering correlation factors please refer to the Dimension Vista System Operator s Guide Rev A 2017-03 or online iGuide (Advanced Functions Section); Method Configuration, Section 9, page 9-25. Process Quality Control (QC) after recalibration and entry of lot specific correlation factors. The correlation factors may create an upward QC shift. Please follow your current laboratory process for adjusting QC ranges. Siemens is providing some examples of CA15-3 lot specific QC values calculated for the three in-date lots of Dimension Vista LOCI CA15-3. CA 15-3 lot specific bottle values for representative sets of QC materials are shown in Table 3. Correlation Factors will remain in the LOCI CA 15-3 Method configuration until removed manually or revised by the user. You may receive future shipments of lots listed in Table 1. Be sure to retain this letter in order to apply correlation factors to CA15-3 lots listed in Table 1. Subsequent lots may require correlation factors. An Alert Card will be included in the subsequent lots with the lot specific correlation factors and lot specific QC. Please review this letter with your Medical Director. If you have any questions, please contact your Siemens Healthineers Customer
Recall details
- Recall number
- Z-1648-2022
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Jul 18, 2022
- Posted by the FDA
- Aug 29, 2022
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 129 units US; 159 units OUS
- Product code
- MOI
- FDA submissions
- K100344
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Austria, France, ¿¿ Germany, Italy, Switzerland.
Product description
Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456
Code information (models, lots, serial numbers)
UDI: 0084276802852621340BB22120610464344840, 0084276802852622049BC23021810464344840, 0084276802852622112BF23042210464344840. Lot Numbers: lot 21340BB Exp Date: 2022-12-06; lot 22049BC (corrected lot number 9/8/2022) Exp Date: 2023-02-18; lot 22112BF Exp Date: 2023-04-20.
Other devices under the same FDA submission
1 device cleared under K100344 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Dimension Vista® Flex® reagent cartridge CA15-3 10464344Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, IMMUNOLOGICAL, ANTIGEN, TUMOR4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.