MaxGuard recallZ-1759-2022 · Carefusion · initiated Aug 1, 2022
Class II Open, Classified All recalls posted in September 2022
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Carefusion or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Administration set, and extension set product labels list expiration year 2027 but the labels should list expiration year 2025. If expired product is used, the integrity of the product may be compromised, and may result in potential leaks and risk of infection to the patient and healthcare worker.
FDA root cause category: Error in labeling
What the company is doing
On 8/1/22 Correction-recall notices were emailed or mailed to customers and distributors who were asked to do the following: 1) Destroy affected devices following your institutions process for destruction. 2) Share this notification with all users within your facility network to ensure all are aware. The correction-recall notice should also be shared with any organization where the potentially affected product has been transferred. 3) Complete and return the response form. In addition distributors were informed that the recalling firm plans to notify all affected customers directly so they were asked to provide a a list of customers who received affected devices. The distributor response form has a check box if the distributor declines to provide a customer list to the recalling firm because they have notified customers about the correction-recall. A copy of the customer communication was provided to the distributor. If customers or distributors require additional assistance they are aske to call 1-844-823-5433, Mon-Fri 8:00 AM - 5:00 PM CT or [email protected]
Recall details
- Recall number
- Z-1759-2022
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 1, 2022
- Posted by the FDA
- Sep 13, 2022
- Recalling firm
- Carefusion
- Quantity
- 100
- Product code
- FPA
- FDA submissions
- K051499
- Distribution
- US: TN, FL, LA, VA, MI, OH, NC, WA, UT, TX, NV, AL, AR, IA, MA, CO, NY, SC, IN, KY, MO, MD, PA, CA, IL, SD, GA, MS, AZ, WV, ID, WY, VT, NM, OR, NJ, OK, KS, MN
Product description
BD MaxGuard Extension Set with 1.2 Micron Filter and Needleless Y-Site, REF: MX9171
Code information (models, lots, serial numbers)
UDI-DI: 10885403237669, Lots: 22029674
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MaxGuard MX9171BD Switzerland Sàrl · Set, administration, intravascular1 recall7 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90707).
- Z-1752-2022 · Open, ClassifiedBD MaxGuard Extension Set with 4-way stopcock, REF: M4058
- Z-1753-2022 · Open, ClassifiedBD MaxGuard Extension Set, REF: ME1069
- Z-1754-2022 · Open, ClassifiedBD MaxGuard Flow Controller Extension Set with Needleless Y-Site, REF: MFS106
- Z-1755-2022 · Open, ClassifiedBD MaxGuard Flow Controller Extension Set with Needleless Y-Site, REF: MFS141
- Z-1756-2022 · Open, ClassifiedBD MaxGuard Drop Administration set with 3 Needleless Y-Sites Removable 3-port Manifold and removable 4-way stopcock, REF: MX4436
- Z-1757-2022 · Open, ClassifiedBD MaxGuard Extension Set with 2 Needleless Y-Site, REF: MX9059
- Z-1758-2022 · Open, ClassifiedBD MaxGuard Extension Set with 0.2 Micron Filter and Needleless Y-Site, REF: MX9166
- Z-1760-2022 · Open, ClassifiedBD MaxGuard 15 Drop Administration Set with 4 Needleless Y-Sites, REF: MX9433
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page