Skip to content

Guider Softip XF recallZ-1582-2022 · Boston Scientific Corporation · initiated Aug 9, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The impacted products were distributed with an MPXF tip curve shape instead of the 40XF tip curve shape for the Guider/40XF/8FR/90CM guide catheter.

FDA root cause category: Process control

What the company is doing

Stryker notified its consignees by issuing an URGENT: MEDICAL DEVICE VOLUNTARY Recall notice on 8/9/2022 by courier. The notice described the problem and the hazard, and requested the consignee take the following actions: "Required Actions 1. Immediately check your internal inventory for affected devices. 2. Segregate the affected units in a secure location for return to Stryker. 3. Circulate this Recall-Removal notice internally to all interested/affected parties. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that Stryker can inform the recipients appropriately. 6. Please inform Stryker of any adverse events concerning the use of the subject devices. 7. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. 8. Return the completed form to your local Stryker Contact. 9. Product Return Information will be provided to you by your designated Sales Representative." UPDATE: On/about 2/13/2023, the amended letter was issued. It identified the lot number (which was in addition to the lot number in the initial recall) and provided the same instructions as were contained in the original notice issued 8/9/2022.

Recall details

Recall number
Z-1582-2022
Status
Open, Classified
Initiated by the company
Aug 9, 2022
Posted by the FDA
Aug 25, 2022
Recalling firm
Boston Scientific Corporation
Quantity
184 devices (UPDATED)
Product code
DQY
FDA submissions
K980453, K010853
Distribution
International distribution in the countries of Greece, France, Germany, Italy, Brazil and South Korea. UPDATE: The devices were also distributed to Russia. There was no U.S. distribution.

Product description

Boston Scientific Guider Softip XF, 40 XF, 8F, Guide Catheter, REF H965100440

Code information (models, lots, serial numbers)

UDI/DI 08714729202486, Lot/Batch 27339850, Use By 2021-05-18, and UPDATE: Lot/Batch 29308869, Use By 4/27/2025.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page