Artis icono biplane- Model No. 11327600 recallZ-1647-2022 · Siemens Medical Solutions USA, Inc · initiated Aug 10, 2022
Class II Open, Classified All recalls posted in August 2022
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
FDA root cause category: Device Design
What the company is doing
On August 10, 2022, the firm notified customers via "URGENT: MEDICAL DEVICE CORRECTION" Letter. Customers were advised that if the error occurs, the user should try to recover normal operation by a shutdown and restart of the system. Siemens will correct the error with a hardware modification via Update Instruction AX028/22/S and AX039/22/S. Siemens' service organization will contact customers shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. If you have any further questions, call 610-219-4834 or email: [email protected].
Recall details
- Recall number
- Z-1647-2022
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 10, 2022
- Posted by the FDA
- Aug 26, 2022
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 1,389 (590 US; 799 OUS) (Total)
- Product code
- OWB
- FDA submissions
- K220432
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (Foreign) to countries of: United Arab Emirates; Armenia; Angola; Argentina; Austria; Australia; Bangladesh; Belgium; Bulgaria; Bahrain; Brazil; Canada; Switzerland; China; Colombia; Costa Rica; Czech Republic; Germany; Denmark; Dominican Republic; Algeria; Estonia; Egypt; Spain; Finland; France; United Kingdom; Greece; Guatemala; Hong Kong; Croatia; Inhouse Company; Indonesia; Ireland; Israel; India; Iran; Italy; Jordan; Japan; South Korea; Kuwait; Kazakhstan; Lebanon; Luxembourg; Monaco; Mongolia; Martinique; Mexico; Malaysia; Netherlands; Norway; New Zealand; Oman; Philippines; Poland; Portugal; Qatar; Serbia; Russian Federation; Saudi Arabia; Sweden; Singapore; Slovenia; Slovakia; El Salvador; Thailand; Turkey; Taiwan; Ukraine; Vietnam and South Africa.
Product description
Artis icono biplane- Model No. 11327600 Artis icono floor- Model No. 11327700
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Artis icono biplane- Model No. 11327600 UDI-DI: 04056869063317 Serial Numbers 180062 180054 180019 180421 180444 180034 180072 180446 180367 180438 180382 180060 180428 180369 180370 180371 180372 180080 180301 180305 180067 180399 180442 180040 180379 180386 180360 180333 180352 180407 180433 180412 180400 180397 180432 180316 180447 180363 180429 180084 180418 180334 180018 180376 180354 180046 180434 180357 180358 180361 180362 180081 180423 180411 180414 180381 180021 180443 180309 180356 180378 180393 180368 180413 180323 180439 180445 180430 180043 180437 180049 180377 180077 180055 180014 180037 180373 ***Updated 9/23/22*** 180466 180465 180464 180047 180463 ***Updated 1/9/23*** 180470 180534 180468 180472 180467 180469 180066 180082 Artis icono floor- Model No. 11327700 UDI-DI: 04056869149325 Serial Numbers 170319 170391 170392 170397 170026 170388 170343 170367 170368 170360 170349 170361 170383 170322 170385 170386 170025 170384 170346 170051 170358 170389 170357 170365 170387 170390 170393 170394 170378 170396 170053 170364 ***Updated 1/9/23*** 170402
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90771).
- Z-1645-2022 · Open, ClassifiedArtis Q floor- Model No. 10848280 Artis Q ceiling- Model No. 10848281 Artis Q biplane- Model No. 10848282 Artis Q zeego- Model No. 10848283 Artis Q.zen floor- Model No. 1084835…
- Z-1646-2022 · Open, ClassifiedArtis pheno- Model No. 10849000
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page