Bentson recallZ-1827-2022 · Cook Incorporated · initiated Aug 16, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
FDA root cause category: Under Investigation by firm
What the company is doing
On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235
Recall details
- Recall number
- Z-1827-2022
- Status
- Open, Classified
- Initiated by the company
- Aug 16, 2022
- Posted by the FDA
- Sep 28, 2022
- Recalling firm
- Cook Incorporated
- Quantity
- 5571 US, 409 OUS
- Product code
- DQX
- FDA submissions
- K171764
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
Product description
Bentson Wire Guide (20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFB-35-80 (GPN G01315) TSFB-35-145 (GPN G00691) TSFB-35-200 (GPN G01095)
Code information (models, lots, serial numbers)
TSFB-35-80 (GPN G01315) UDI-DI 00827002270411 Lots 14751376 Expiration Date 25-05-2027 TSFB-35-145 (GPN G00691) UDI-DI 00827002006911 Lots Expiration Dates 14813687 27-06-2027 14831302 07-07-2027 14831303 07-07-2027 14831304 07-07-2027 14831305 07-07-2027 14831307 07-07-2027 14831308 07-07-2027 14831309 07-07-2027 14831310 07-07-2027 14831311 07-07-2027 14831313 07-07-2027 14842394 14-07-2027 14842395 14-07-2027 14842396 14-07-2027 14842397 14-07-2027 14842398 14-07-2027 14842399 14-07-2027 14842400 14-07-2027 14842401 14-07-2027 TSFB-35-200 (GPN G01095) UDI-DI 00827002010956 Lot/Expiration Date: 14816668 28-06-2027
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Bentson G27041Cook Incorporated · WIRE, GUIDE, CATHETER2 recalls0 reports
- Bentson G00691Cook Incorporated · WIRE, GUIDE, CATHETER1 recall26 reports
- Bentson G01095Cook Incorporated · WIRE, GUIDE, CATHETER1 recall1 report
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90793).
- Z-1824-2022 · Open, ClassifiedAmplatz extra Stiff Wire Guide, Stiff Guidewire (OUS ONLY); Reference Part Number: THSF-35-145-AES-SGH (GPN G06801) THSF-38-80-AES-SGH (GPN G06874)
- Z-1825-2022 · Open, ClassifiedBentson Plus Wire Guide (Extra Body/20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFBP-35-180 (GPN G09782) TSFBP-35-145 (GPN G09735)
- Z-1826-2022 · Open, ClassifiedBentson Wire Guide, Guidewire (OUS ONLY); Reference Part Number: TSMB-35-145 (GPN G27041)
- Z-1828-2022 · Open, ClassifiedCoons Interventional Wire Guide, Guidewire; Reference Part Number/GPN: THSF-35-145-COONS G02356 THSF-38-145-COONS G02324 THSF-35-180-COONS G02621 THSF-35-80-COONS G12268
- Z-1829-2022 · Open, ClassifiedCope Mandril Wire Guide (Nitinol), Nitinol Guidewire; Reference Part Number/GPN: PMG-18SP-60-COPE-NT-ST G09395 PMG-18SP-100-COPE-NT G08687 PMG-18SP-60-COPE-NT G08427
- Z-1830-2022 · Open, ClassifiedCope Mandril Wire Guide (Stainless Steel), Stainless Steel Guidewire Reference Part Number/GPN PMG-18SP-60-COPE G05183 PMG-18SP-100-COPE G05822 PMG-18SP-125-COPE G10077
- Z-1831-2022 · Open, ClassifiedDouble Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN TDOC-35-145-0-3 G00362 C-TDOC-18-50-0-2 G05146 TDOC-35-50-0-3 G00364 C-DOC-21-40-0-2 G02431 C-TDOC-21-50-0-2 …
- Z-1832-2022 · Open, ClassifiedFixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476
- Z-1833-2022 · Open, ClassifiedFixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference Part Number/GPN: TSCF-25-180-3 G01602 TSCF-35-180-3 G00902 TSCF-21-260-3 G00455 THSCF-25-260-3 G02384 TSCF-25-145-3 G00…
- Z-1834-2022 · Open, ClassifiedFixed Core Wire Guide (Straight), Guidewire Reference Part Number/GPN: C-SF-18-30 G02261 C-SF-25-40 G01937 TSF-18-260 G00590 C-SF-25-50 G01938 TSF-35-180 G00652 C-SF-18-40 G02255 T…
- Z-1835-2022 · Open, ClassifiedHeavy Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN: THDOC-35-60-0-3 G11949 C-HDOC-18-40-0-2 G09201 C-THDOC-18-40-0-2-HTS G26925
- Z-1836-2022 · Open, ClassifiedNewton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701 TSCFNA-35-145-3 G00561
- Z-1837-2022 · Open, ClassifiedRoadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5 G34132 RPC-038145-0-5 G34129 RFSPC-03814…
- Z-1838-2022 · Open, ClassifiedRoadrunner PC Wire Guide (Nimble), Guidewire Reference Part Number/GPN: RFSPC-35-145 G09607
- Z-1839-2022 · Open, ClassifiedRoadrunner PC Wire Guide (Nimble Floppy), Guidewire RFPC-35-145 G07914
- Z-1840-2022 · Open, ClassifiedRoadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866
- Z-1841-2022 · Open, ClassifiedRoadrunner PC Wire Guide (The Firm), Guidewire Reference Part Number/GPN RPC-35-145 G06979 RPC-35-80 G09534
- Z-1842-2022 · Open, ClassifiedRosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623 THSCF-35-80-1.5-ROSEN G01774 THSCF-35-180-1.5-ROSEN G01264 THSCF-35-145-1.5-ROSEN G01261
- Z-1843-2022 · Open, ClassifiedTorq-Flex Wire Guide Australian Modification, Guidewire Reference Part Number/GPN STF-18-40-AUST G07304
- Z-1844-2022 · Open, ClassifiedVinyl Connecting Tube, Connecting Tube Reference Part Number/GPN CTU14.0-30-ST G02278 CTU14.0-40-ST G02327 CTU14.0-30 G02791 CTU14.0-50-ST G02464 C-CTU14.0-30-ST G02898
- Z-1845-2022 · Open, ClassifiedLunderquist -Ring Torque Wire Guide, Guidewire Reference Part Number/GPN THSF-35-125-THG G27031 THSF-35-145-THG G27033
- Z-1846-2022 · Open, ClassifiedMovable Core Wire Guide (Tapered Tefcor) (OUS ONLY), Guidewire Reference Part Number/GPN TCMTNA-35-145-3 G01678
- Z-1847-2022 · Open, ClassifiedMovable Core Wire Guide (Tefcor), Guidewire (OUS ONLY) Reference Part Number/GPN TMT-35-145 G01677
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page