Skip to content

Roadrunner recallZ-1837-2022 · Cook Incorporated · initiated Aug 16, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cook Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

FDA root cause category: Under Investigation by firm

What the company is doing

On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235

Recall details

Recall number
Z-1837-2022
Status
Open, Classified
Initiated by the company
Aug 16, 2022
Posted by the FDA
Sep 28, 2022
Recalling firm
Cook Incorporated
Quantity
56 US; 1033 OUS
Product code
OCY
FDA submissions
K082536
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.

Product description

Roadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5 G34132 RPC-038145-0-5 G34129 RFSPC-038145-0 G17539 RFSPC-035145 G17540 RPC-038145-0 G18155 RPC-035145-0 G18153 RFSPC-035145-0 G17542

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

RPC-035145-0-5 G34131 UDI-DI: 00827002341319 Lot/Expiration Date: 14769270X 03-06-2025 14802181 20-06-2025 14816224 28-06-2025 14818488 29-06-2025 14818489 29-06-2025 14828627 06-07-2025 14828629 06-07-2025 RPC-035145 G18154 UDI-DI: 00827002181540 Lot/Expiration Date: 14770863 06-06-2025 14770864 06-06-2025 14813490 27-06-2025 14816227 28-06-2025 NS14770866 06-06-2025 NS14804982 21-06-2025 RPC-035145-5 G34132 UDI-DI: 00827002341326 Lot/Expiration Date: 14770872 06-06-2025 14810022 23-06-2025 14831440 07-07-2025 RPC-038145-0-5 G34129 UDI-DI: 00827002341296 Lot/Expiration Date: 14804972 21-06-2025 14809971 23-06-2025 RFSPC-038145-0 G17539 UDI-DI: 00827002175396 Lot/Expiration Date: 14809157 23-06-2025 14837048 12-07-2025 RFSPC-035145 G17540 UDI-DI: 00827002175402 Lot/Expiration Date: 14818484 29-06-2025 14818485 29-06-2025 14818490 29-06-2025 14840310 13-07-2025 RPC-038145-0 G18155 UDI-DI: 00827002181557 Lot/Expiration Date: 14819269 29-06-2025 14823406 01-07-2025 14832976 08-07-2025 NS14828649 06-07-2025 RPC-035145-0 G18153 UDI-DI: 00827002181533 Lot/Expiration Date: 14828625 06-07-2025 14828630 06-07-2025 14828635 06-07-2025 14828636 06-07-2025 14828637 06-07-2025 14828640 06-07-2025 14828641 06-07-2025 14841469 14-07-2025 NS14813479 27-06-2025 NS14818486 29-06-2025 RFSPC-035145-0 G17542 UDI-DI: 00827002175426 Lot/Expiration Date: 14844152 15-07-2025

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90793).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page