Hu-Friedy K360 Scalpel Handle, 360 Degree Scalpel Handle… recallZ-0079-2023 · Steris Corporation · initiated Sep 2, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The scalpel holder may separate from the handle for certain lots. If the tip falls off during use, it could fall into the patient's mouth which may result in a cut to the patient's mouth and also presents a choking hazard.

FDA root cause category: Process design

What the company is doing

Starting September 2, 2022, Hu-Friedy notified all affected customers and distributors via Urgent Medical Device Recall letters. Customers were instructed to immediately destroy all on-hand inventory of the affected lots of K360 Scalpel Handles. Distributors should identify and notify any of their customers who have received further distributed product. Hu-Friedy will issue credit upon receipt of completed response forms.

Recall details

Recall number
Z-0079-2023
Status
Terminated
Initiated by the company
Sep 2, 2022
Posted by the FDA
Oct 12, 2022
Recalling firm
Steris Corporation
Quantity
5 US; 62 OUS
Product code
GDZ
Distribution
Distribution to US states of IL, MA, NC, SC, and TN, Canada, China, Japan, Malaysia, Turkey, Great Britain, and India.

Product description

Hu-Friedy K360 Scalpel Handle, 360 Degree Scalpel Handle, REF K360

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 10889950014193 Lots 0322, 0521, 0522, and 0621