Medical device recalls: October 2022
The FDA posted 153 medical device recall records in October 2022, from 51 recall events by 44 companies. 13 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- MEDLINE INDUSTRIES, LP - Northfield · 3 events, 7 records
- Baxter Healthcare Corporation · 2 events, 10 records, 5 Class I
- Synthes (USA) Products LLC · 2 events, 5 records
- Wright Medical Technology, Inc. · 2 events, 4 records
- Boston Scientific Corporation · 2 events, 3 records
- Siemens Healthcare Diagnostics, Inc. · 2 events, 3 records
- Hobbs Medical, Inc. · 1 event, 61 records
- Medicina Uk Ltd · 1 event, 8 records
FDA root cause
- Process control · 72
- Under Investigation by firm · 31
- Error in labeling · 11
- Device Design · 9
- Nonconforming Material/Component · 8
- Process design · 4
- Software design · 2
- Process change control · 2
Class I recalls (13)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0141-2023 · Baxter Healthcare Corporation · posted Oct 31, 2022 · Open, Classified
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
- Z-0142-2023 · Baxter Healthcare Corporation · posted Oct 31, 2022 · Open, Classified
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
- Z-0143-2023 · Baxter Healthcare Corporation · posted Oct 31, 2022 · Open, Classified
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
- Z-0144-2023 · Baxter Healthcare Corporation · posted Oct 31, 2022 · Open, Classified
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
- Z-0145-2023 · Baxter Healthcare Corporation · posted Oct 31, 2022 · Open, Classified
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
- Z-0072-2023 · Teleflex LLC · posted Oct 14, 2022 · Open, Classified
Incidents of device splitting or detaching during use
- Z-0073-2023 · Teleflex LLC · posted Oct 14, 2022 · Open, Classified
Incidents of device splitting or detaching during use
- Z-0074-2023 · Teleflex LLC · posted Oct 14, 2022 · Open, Classified
Incidents of device splitting or detaching during use
- Z-0075-2023 · Teleflex LLC · posted Oct 14, 2022 · Open, Classified
Incidents of device splitting or detaching during use
- Z-0076-2023 · Teleflex LLC · posted Oct 14, 2022 · Open, Classified
Incidents of device splitting or detaching during use
- Z-0077-2023 · Teleflex LLC · posted Oct 14, 2022 · Open, Classified
Incidents of device splitting or detaching during use
- Z-0001-2023 · Jiangsu Well Biotech Co.,Ltd. · posted Oct 8, 2022 · Open, Classified
Distribution of COVID-19 Ag Rapid Test kits in the U.S. without an Emergency Use Authorization, or a Pre-Market Approval or Clearance.
- Z-0002-2023 · Fresenius Kabi USA, LLC · posted Oct 6, 2022 · Terminated
The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or…
Class II recalls (139)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0150-2023 · Ethicon, Inc. · posted Oct 31, 2022 · Open, Classified
Packages may contain a PS-1 cutting edge needle and not the intended SH taper point needle may result in more tissue trauma than is anticipated for taper point needles, which may necessitate repair
- Z-0151-2023 · Boston Scientific Corporation · posted Oct 31, 2022 · Open, Classified
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
- Z-0152-2023 · Boston Scientific Corporation · posted Oct 31, 2022 · Open, Classified
There is an incorrect manufacturing date/timestamp within the software which causes an inaccurate display of battery capacity.
- Z-0153-2023 · Belmont Instrument Corporation · posted Oct 31, 2022 · Open, Classified
Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The issue may potentially lead to delayed treatment and a failure to deliver the prescribed thermoregulation therapy. To…
- Z-0146-2023 · Edwards Lifesciences, LLC · posted Oct 28, 2022 · Open, Classified
Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.
- Z-0147-2023 · Edwards Lifesciences, LLC · posted Oct 28, 2022 · Open, Classified
Due to an increase of confirmed latex deterioration complaints for arterial embolectomy catheters packaged in pouches.
- Z-0148-2023 · SenTec AG · posted Oct 28, 2022 · Terminated
Weak skin adhesion.
- Z-0149-2023 · Boston Scientific Corporation · posted Oct 28, 2022 · Open, Classified
Consoles were shipped without an instruction for use (IFU) or with the incorrect IFU.
- Z-0137-2023 · Merit Medical Systems, Inc. · posted Oct 27, 2022 · Open, Classified
Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.
- Z-0138-2023 · Merit Medical Systems, Inc. · posted Oct 27, 2022 · Open, Classified
Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.
- Z-0139-2023 · Wright Medical Technology, Inc. · posted Oct 27, 2022 · Terminated
The tibial tray lock detail is oversized (larger than specification).
- Z-0140-2023 · Wright Medical Technology, Inc. · posted Oct 27, 2022 · Terminated
The tibial tray lock detail is oversized (larger than specification).
- Z-0134-2023 · Luminex Corporation · posted Oct 26, 2022 · Open, Classified
A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result.
- Z-0136-2023 · W L Gore & Associates, Inc. · posted Oct 26, 2022 · Open, Classified
Due to manufacturing records (Release Test Results) indicating "Failed".
- Z-0132-2023 · Angiodynamics, Inc. · posted Oct 25, 2022 · Open, Classified
Printed Instructions for Use (IFU) not shipped with device. The Instructions for Use contain the Warnings, Precautions, Contraindications, and Procedures for the Preparation, Implantation, and Post Operative Care of the implantable port, failure to follow…
- Z-0133-2023 · Intuitive Surgical, Inc. · posted Oct 25, 2022 · Open, Classified
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force…
- Z-0127-2023 · ConvaTec, Inc · posted Oct 24, 2022 · Open, Classified
impaired functionality of the release liner
- Z-0128-2023 · ConvaTec, Inc · posted Oct 24, 2022 · Open, Classified
impaired functionality of the release liner
- Z-0129-2023 · ConvaTec, Inc · posted Oct 24, 2022 · Open, Classified
impaired functionality of the release liner
- Z-0130-2023 · ConvaTec, Inc · posted Oct 24, 2022 · Open, Classified
impaired functionality of the release liner
- Z-0131-2023 · Datascope Corporation · posted Oct 24, 2022 · Open, Classified
The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy. Therapy delays may impact those patients…
- Z-0118-2023 · ZOLL Circulation, Inc. · posted Oct 21, 2022 · Open, Classified
Due to increase in Li-Ion Battery failures
- Z-0119-2023 · Siemens Medical Solutions USA, Inc · posted Oct 21, 2022 · Open, Classified
Siemens has become aware of three potential software issues with ARTIS pheno systems with software version VE10B. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be continued on an…
- Z-0120-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 21, 2022 · Open, Classified
Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration date on a component with the kit. CHG Solution with the Pre-Op Changing Bathing Kit and Bactec bottle plastic both have an expiration date that does not match…
- Z-0121-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 21, 2022 · Open, Classified
Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration date on a component with the kit. CHG Solution with the Pre-Op Changing Bathing Kit and Bactec bottle plastic both have an expiration date that does not match…
- Z-0122-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 21, 2022 · Open, Classified
Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration date on a component with the kit. CHG Solution with the Pre-Op Changing Bathing Kit and Bactec bottle plastic both have an expiration date that does not match…
- Z-0123-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 21, 2022 · Open, Classified
Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration date on a component with the kit. CHG Solution with the Pre-Op Changing Bathing Kit and Bactec bottle plastic both have an expiration date that does not match…
- Z-0124-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 21, 2022 · Open, Classified
Adult Blood Culture Collect kit and Pre-Op Changing Bathing Kit contain an incorrect expiration date on a component with the kit. CHG Solution with the Pre-Op Changing Bathing Kit and Bactec bottle plastic both have an expiration date that does not match…
- Z-0125-2023 · Encore Medical, LP · posted Oct 21, 2022 · Open, Classified
Hip liner implant boxes may contain knee inserts and knee insert boxes may contain hip liner implants, which may lead to implants not being available for surgeries, which may cause delays in surgery.
- Z-0126-2023 · Encore Medical, LP · posted Oct 21, 2022 · Open, Classified
Hip liner implant boxes may contain knee inserts and knee insert boxes may contain hip liner implants, which may lead to implants not being available for surgeries, which may cause delays in surgery.
- Optima Coil SystemClass IIZ-0103-2023 · Balt USA, LLC · posted Oct 20, 2022 · Open, Classified
Product pouch label does not match up with carton label
- Z-0104-2023 · Siemens Healthcare Diagnostics, Inc. · posted Oct 20, 2022 · Open, Classified
Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting…
- Z-0105-2023 · Siemens Healthcare Diagnostics, Inc. · posted Oct 20, 2022 · Open, Classified
Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting…
- Z-0106-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 20, 2022 · Terminated
One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.
- Z-0107-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0108-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0109-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Medicina Oral Tip Syringe 1ml ↗Class IIZ-0110-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0111-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0112-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0113-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0114-2023 · Medicina Uk Ltd · posted Oct 20, 2022 · Open, Classified
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
- Z-0115-2023 · Abbott Vascular · posted Oct 20, 2022 · Open, Classified
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to…
- Z-0116-2023 · Abbott Vascular · posted Oct 20, 2022 · Open, Classified
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to…
- Z-0117-2023 · Abbott Vascular · posted Oct 20, 2022 · Open, Classified
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to…
- Z-0102-2023 · ARROW INTERNATIONAL Inc. · posted Oct 19, 2022 · Terminated
Mislabeling: the corrugate label, lidstock, banner card, and peel sticker accompanying the affected product incorrectly displays a double lumen 5.5 French catheter (and related priming volumes and kit contents for product code ASK-41552-JHVA1), instead of…
- Z-0098-2023 · Synthes (USA) Products LLC · posted Oct 18, 2022 · Open, Classified
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone…
- Z-0099-2023 · Synthes (USA) Products LLC · posted Oct 18, 2022 · Open, Classified
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone…
- Z-0100-2023 · Synthes (USA) Products LLC · posted Oct 18, 2022 · Open, Classified
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone…
- Z-0101-2023 · Synthes (USA) Products LLC · posted Oct 18, 2022 · Open, Classified
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone…
- Z-0096-2023 · Intersurgical Inc · posted Oct 17, 2022 · Terminated
Printing error on the device, size 3 Solus Standard laryngeal mask airway has been incorrectly marked with size 4 information (includes the number 4 as a size identifier, patient weight and maximum inflation volume on the airway tube and the marking #4 on…
- Z-0097-2023 · Roche Molecular Systems, Inc. · posted Oct 17, 2022 · Terminated
Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.
- Z-0089-2023 · Siemens Healthcare Diagnostics, Inc. · posted Oct 14, 2022 · Open, Classified
Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be…
- Z-0090-2023 · B. Braun Medical, Inc. · posted Oct 14, 2022 · Open, Classified
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both…
- Z-0091-2023 · B. Braun Medical, Inc. · posted Oct 14, 2022 · Open, Classified
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both…
- Z-0092-2023 · B. Braun Medical, Inc. · posted Oct 14, 2022 · Open, Classified
The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may…
- Z-0093-2023 · Siemens Healthcare Diagnostics Inc · posted Oct 14, 2022 · Open, Classified
N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
- Z-0094-2023 · Siemens Healthcare Diagnostics Inc · posted Oct 14, 2022 · Open, Classified
N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
- Z-0095-2023 · Becton Dickinson & Company · posted Oct 14, 2022 · Open, Classified
The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in methanol testing.
- Z-0087-2023 · Shanghai United Imaging Healthcare Co., Ltd. · posted Oct 13, 2022 · Open, Classified
The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the…
- VENCLOSE, VENCLOSE PROCEDURE PACK, 7VM, REF VC-PPH-67A. Cardiovascular procedure convenience kit.Class IIZ-0088-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Oct 13, 2022 · Terminated
Impacted Venclose Procedure Packs contain a 12cm introducer/sheath component, either component #128626 - SET INTRO 7FX12CM FITS 0.018 or component #137339 - SET INTRO 6.5FX12CM FITS 0.018, instead of a 7cm introducer/sheath, component #137340 - SET INTRO…
- Z-0078-2023 · Olympus Corporation of the Americas · posted Oct 12, 2022 · Open, Classified
Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with…
- Z-0079-2023 · Steris Corporation · posted Oct 12, 2022 · Terminated
The scalpel holder may separate from the handle for certain lots. If the tip falls off during use, it could fall into the patient's mouth which may result in a cut to the patient's mouth and also presents a choking hazard.
- Z-0080-2023 · LUMINOSTICS, INC dba CLIP HEALTH · posted Oct 12, 2022 · Open, Classified
COVID test kit shelf life was periodically extended, as real-time stability test timepoints passed, without FDA concurrence.
- Z-0081-2023 · Baxter Healthcare Corporation · posted Oct 12, 2022 · Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Z-0082-2023 · Baxter Healthcare Corporation · posted Oct 12, 2022 · Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Z-0083-2023 · Baxter Healthcare Corporation · posted Oct 12, 2022 · Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Z-0084-2023 · Baxter Healthcare Corporation · posted Oct 12, 2022 · Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Z-0085-2023 · Baxter Healthcare Corporation · posted Oct 12, 2022 · Open, Classified
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
- Z-0086-2023 · Johnson & Johnson Surgical Vision, Inc. · posted Oct 12, 2022 · Open, Classified
As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.
- COVID Test Kit Nonsterile ↗Class IIZ-0071-2023 · American Contract Systems, Inc. · posted Oct 8, 2022 · Open, Classified
Kit incorrectly labeled for use with COVID testing and should not be used with any SARS-Co-V2 device.
- Z-0010-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0011-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0012-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Medical Cytology Brush ↗Class IIZ-0013-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Medical Cytology Brush ↗Class IIZ-0014-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0015-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0016-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0017-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0018-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Scope Introducer ↗Class IIZ-0019-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0020-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0021-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0022-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0023-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0024-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0025-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0026-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0027-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0028-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0029-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0030-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0031-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0032-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Helical Retrieval Basket ↗Class IIZ-0033-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Medical Grasping Forceps ↗Class IIZ-0034-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Medical Grasping Forceps ↗Class IIZ-0035-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Hobbs Retrieval Forceps ↗Class IIZ-0036-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0037-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0038-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0039-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0040-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0041-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0042-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0043-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0044-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0045-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0046-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0047-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0048-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0049-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0050-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0051-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0052-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0053-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0054-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0055-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0056-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0057-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0058-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0059-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0060-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0061-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0062-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0063-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0064-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Freeman Pancreatic Flexi-Stent ↗Class IIZ-0065-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0066-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0067-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0068-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0069-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0070-2023 · Hobbs Medical, Inc. · posted Oct 7, 2022 · Open, Classified
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending…
- Z-0008-2023 · Synthes (USA) Products LLC · posted Oct 6, 2022 · Open, Classified
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation.…
- Z-0009-2023 · B Braun Medical Inc · posted Oct 6, 2022 · Open, Classified
Potential for leakage at the catheter hub.
- Z-0003-2023 · In My Bathroom LLC · posted Oct 5, 2022 · Open, Classified
Excessive ultraviolet-C radiation
- Z-0007-2023 · Zimmer, Inc. · posted Oct 5, 2022 · Open, Classified
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection…
- Z-0004-2023 · Philips North America Llc · posted Oct 4, 2022 · Open, Classified
Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.
- Z-0005-2023 · Wright Medical Technology, Inc. · posted Oct 4, 2022 · Open, Classified
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
- Z-0006-2023 · Wright Medical Technology, Inc. · posted Oct 4, 2022 · Open, Classified
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0135-2023 · LumiraDx · posted Oct 26, 2022 · Terminated
Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2…