Micro-X Rover Mobile X-Ray System, Mxu-Rv19 recallZ-1786-2022 · Micro-X Ltd. · initiated Aug 31, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Micro-X Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

FDA root cause category: Process control

What the company is doing

On or about 09/31/2022, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via email/telephone to customers informing them that during an internal data analysis on 08/18/2022, Micro-X Ltd identified software calibration errors with the product equipped with a Dose Area Product (DAP) meter. This resulted in the product displaying incorrect DAP meter values. The DICOM image produced by the system will report a DAP measurement in microgray/square meter that is 8 to 10 times lower than the actual dose delivered. Customers are instructed to: 1. Inspect their stock immediately and place the Recall Notification Letter on the affected product and quarantine affected product on hand to prevent further use. 2. Contact the nominated Service representative to arrange the re-calibration of the DAP Meter. Email: [email protected] or Phone 1-206-249-8764. 3.Previous version of Service Documentation (ROV208-2 Acceptance and Compliance Testing Instructions) which includes inaccurate DAP calibration processes are to be removed from use. 4. Complete and return Acknowledgement and Receipt Form immediately to Micro-X LTD in Australia: [email protected] and copy [email protected] 5. Ensure relevant staff members are informed of the Recall and if affected units have been supplied/transferred to another facility, inform them immediately. For question or further assistance, contact Regulatory Affairs Manager at [email protected]

Recall details

Recall number
Z-1786-2022
Status
Open, Classified
Initiated by the company
Aug 31, 2022
Posted by the FDA
Sep 23, 2022
Recalling firm
Micro-X Ltd.
Quantity
42 units
Product code
IZL
FDA submissions
K211423
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.

Product description

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model Number: MXU-RV19 UDI Codes: (01)09357123000013(11)210326(21)00222 (01)09357123000013(11)210628(21)00234 (01)09357123000013(11)210629(21)00235 (01)09357123000013(11)210629(21)00236 (01)09357123000013(11)210630(21)00237 (01)09357123000013(11)210708(21)00238 (01)09357123000013(11)210721(21)00240 (01)09357123000013(11)210910(21)00261 (01)09357123000013(11)210913(21)00262 (01)09357123000013(11)210914(21)00263 (01)09357123000013(11)210921(21)00264 (01)09357123000013(11)210922(21)00265 (01)09357123000013(11)210923(21)00266 (01)09357123000013(11)210927(21)00267 (01)09357123000013(11)211006(21)00268 (01)09357123000013(11)211008(21)00269 (01)09357123000013(11)211011(21)00270 (01)09357123000013(11)211011(21)00271 (01)09357123000013(11)211012(21)00272 (01)09357123000013(11)211013(21)00273 (01)09357123000013(11)211013(21)00274 (01)09357123000013(11)211210(21)00288 (01)09357123000013(11)211213(21)00289 (01)09357123000013(11)211213(21)00290 (01)09357123000013(11)220329(21)00305 (01)09357123000013(11)220330(21)00306 (01)09357123000013(11)220331(21)00307 (01)09357123000013(11)220401(21)00308 (01)09357123000013(11)220404(21)00309 (01)09357123000013(11)220405(21)00310 (01)09357123000013(11)220406(21)00311 (01)09357123000013(11)220404(21)00312 Serial Numbers: 222 234 235 236 237 238 240 261 262 263 264 265 266 267 268 269 270 271 272 273 274 288 289 290 305 306 307 308 309 310 311 312

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K211423 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page