Micro-X Rover (REF MXU-RV35)Recalls & reported problems

Micro-X Limited · The device is designed to perform radiographic x-ray examinations on pediatric and adult patients, in all patient treatment areas · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

7 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
7

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

7 FDA recall records name this device's identifier, 7 still open.

3 other recalls cover devices cleared under the same FDA submission (K211423); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Mar 27, 2026Class IIOpen, Classified

    X-Ray generator may malfunction resulting in the X-Ray being inoperable.

    Z-1918-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

  2. Apr 4, 2025Class IIOpen, Classified

    Potential for internal fastening mechanism within generator of mobile x-Ray to fail.

    Z-1670-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. Mar 28, 2025Class IIOpen, Classified

    Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.

    Z-1578-2025 · Root cause: Device Design · Recall details · FDA record ↗

  4. Mar 28, 2025Class IIOpen, Classified

    Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.

    Z-1579-2025 · Root cause: Device Design · Recall details · FDA record ↗

  5. Oct 22, 2024Class IIOpen, Classified

    Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.

    Z-0580-2025 · Root cause: Environmental control · Recall details · FDA record ↗

  6. Nov 28, 2023Class IIOpen, Classified

    Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.

    Z-0665-2025 · Root cause: Device Design · Recall details · FDA record ↗

  7. May 31, 2023Class IIOpen, Classified

    Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

    Z-0664-2025 · Root cause: Software design · Recall details · FDA record ↗

FDA clearance and classification

Cleared through Special 510(k) K211423 on May 21, 2021 (decision: Substantially Equivalent). Submitted by Micro-X Limited as “Rover”. The FDA review took 14 days. FDA record ↗

The FDA regulates product code MQB (solid state x-ray imager (flat panel/digital imager)) under 21 CFR 892.1680 as a Class II device. 21 CFR 892.1680 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K211423. One submission often covers many models, so check the product and lot details in each record.

  • Z-1669-2025 · Apr 4, 2025 · Open, Classified
    Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71, MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system th
  • Z-0666-2025 · Oct 23, 2024 · Open, Classified
    MICRO-X Rover Mobile X-ray System, # MXU-RV19
  • Z-1786-2022 · Aug 31, 2022 · Open, Classified
    MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Common questions about the Micro-X Rover

Has the Micro-X Rover been recalled?

Yes. 7 FDA recall records name this device's identifier: 7 still open. The most recent, Z-1918-2026 (Class II), began on Mar 27, 2026: X-Ray generator may malfunction resulting in the X-Ray being inoperable. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the Micro-X Rover FDA approved?

It was cleared through Special 510(k) K211423 on May 21, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Micro-X Rover, and is it still sold?

The FDA's device database lists Micro-X Limited as the company responsible for it (brand Micro-X, model Rover). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Solid State X-Ray Imager (Flat Panel/Digital Imager) devices

All Solid State X-Ray Imager (Flat Panel/Digital Imager) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page