Prismaflex ST Set: ST60 (product code 107643), ST100… recallZ-0084-2023 · Baxter Healthcare Corporation · initiated Sep 14, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
FDA root cause category: Error in labeling
What the company is doing
Baxter sent an Urgent Medical Device Correction notice to its direct consignees by USPS on 09/14/2022. The notice explained the Estonian translation issue and requested that the following actions be taken by consumers: 1. Use of the Prismaflex HF20 set should be restricted to patients with a body weight greater than 8kg (18lb). Use of the Prismaflex M60 and ST60 sets should be restricted to patients with a body weight greater than 11kg (24lb). Use of the Prismaflex M100, ST100, M150, ST150, HF1000 and HF1400 sets should be restricted to patients with a body weight greater than 30kg (66lb). 2. Customers who are not referring to the Estonian IFU should continue to follow the IFU instructions in their language. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com/. 4. If you purchased this product from a distributor, please note that responding via the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal.
Recall details
- Recall number
- Z-0084-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 14, 2022
- Posted by the FDA
- Oct 12, 2022
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 1,068,560 devices
- Product code
- KDI
- FDA submissions
- K212216
- Distribution
- US Nationwide
Product description
Prismaflex ST Set: ST60 (product code 107643), ST100 (product code 107636), ST150 (product code 107640). For use in providing continuous fluid management and renal replacement therapies.
Code information (models, lots, serial numbers)
UDI: 07332414075682 (product code 107643), 07332414075613 (product code 107636), 07332414075651 (product code 107640); All Lot Codes
Other devices under the same FDA submission
3 devices cleared under K212216 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PRISMAFLEX SET 107640USGambro Industries · Dialyzer, high permeability with or without sealed dialysate system1 recall26 reports
- PRISMAFLEX SET 107636USGambro Industries · Dialyzer, high permeability with or without sealed dialysate system1 recall9 reports
- PRISMAFLEX SET 107643USGambro Industries · Dialyzer, high permeability with or without sealed dialysate system1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 90841).
- Z-0081-2023 · Open, ClassifiedPrismaflex M Set: M60 (product code 106696), M100 (product code 106697). For use in providing continuous fluid management and renal replacement therapies.
- Z-0082-2023 · Open, ClassifiedPrismaflex M Set: M150 (product code 109990). For use in providing continuous fluid management and renal replacement therapies.
- Z-0083-2023 · Open, ClassifiedPrismaflex HF Set: HF1000 (product code 107140), HF1400 (product code 107142). For use in providing continuous fluid management and renal replacement therapies.
- Z-0085-2023 · Open, ClassifiedPrismaflex HF20 Set (product code 109841). For use in providing continuous fluid management and renal replacement therapies.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page