Skip to content

Prismaflex ST Set: ST60 (product code 107643), ST100… recallZ-0084-2023 · Baxter Healthcare Corporation · initiated Sep 14, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.

FDA root cause category: Error in labeling

What the company is doing

Baxter sent an Urgent Medical Device Correction notice to its direct consignees by USPS on 09/14/2022. The notice explained the Estonian translation issue and requested that the following actions be taken by consumers: 1. Use of the Prismaflex HF20 set should be restricted to patients with a body weight greater than 8kg (18lb). Use of the Prismaflex M60 and ST60 sets should be restricted to patients with a body weight greater than 11kg (24lb). Use of the Prismaflex M100, ST100, M150, ST150, HF1000 and HF1400 sets should be restricted to patients with a body weight greater than 30kg (66lb). 2. Customers who are not referring to the Estonian IFU should continue to follow the IFU instructions in their language. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com/. 4. If you purchased this product from a distributor, please note that responding via the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal.

Recall details

Recall number
Z-0084-2023
Status
Open, Classified
Initiated by the company
Sep 14, 2022
Posted by the FDA
Oct 12, 2022
Recalling firm
Baxter Healthcare Corporation
Quantity
1,068,560 devices
Product code
KDI
FDA submissions
K212216
Distribution
US Nationwide

Product description

Prismaflex ST Set: ST60 (product code 107643), ST100 (product code 107636), ST150 (product code 107640). For use in providing continuous fluid management and renal replacement therapies.

Code information (models, lots, serial numbers)

UDI: 07332414075682 (product code 107643), 07332414075613 (product code 107636), 07332414075651 (product code 107640); All Lot Codes

Other devices under the same FDA submission

3 devices cleared under K212216 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90841).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page