Skip to content

Extension Set recallZ-0091-2023 · B. Braun Medical, Inc. · initiated Sep 15, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from B. Braun Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

FDA root cause category: Process design

What the company is doing

On September 15, 2022, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were instructed to cease use and quarantine the product subject to recall. Once customers have returned the completed acknowledgement form to B. Braun Medical Inc., a Customer Support representative will contact the customer with instructions on returning impacted cases. The firm will provide credit and/or replacement of the product based on the customer's needs. For questions, please contact BBMI's Postmarket Surveillance Department at 1-833-425-1464.

Recall details

Recall number
Z-0091-2023
Status
Open, Classified
Initiated by the company
Sep 15, 2022
Posted by the FDA
Oct 14, 2022
Recalling firm
B. Braun Medical, Inc.
Quantity
354,200 US; 1,800 OUS
Product code
FPA
FDA submissions
K942988, K954123, K083723
Distribution
Distribution US nationwide, Canada and Saudi Arabia.

Product description

Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 04046964540943 Lots/Expiration Dates 61705083 31-Oct-22 61705124 31-Oct-22 61706055 31-Oct-22 61714103 31-Dec-22 61719696 31-Jan-23 61732456 31-Mar-23 61736580 31-May-23 61739427 31-May-23 61744671 31-Jul-23 61747367 31-Aug-23 61750206 31-Aug-23 61752658 30-Sep-23 61761422 30-Nov-23 61776447 30-Apr-24 61780880 30-Apr-24 61802320 30-Sep-24 61806067 31-Oct-24 61807459 31-Oct-24 61822282 28-Feb-25

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90850).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page