Intersurgical Solus Standard, laryngeal mask airway, size… recallZ-0096-2023 · Intersurgical Inc · initiated Aug 29, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intersurgical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Printing error on the device, size 3 Solus Standard laryngeal mask airway has been incorrectly marked with size 4 information (includes the number 4 as a size identifier, patient weight and maximum inflation volume on the airway tube and the marking #4 on the inflation line pilot balloon) may pose health risk to the patient

FDA root cause category: Under Investigation by firm

What the company is doing

Intersurgical Inc. issued Urgent Medical Device Recall Letter via email on 08/29/2022 and via certified first-class mail to Distributors. Letter states reason for recall, health risk and action to take: The recall extended to the user level. Recall Instructions: i. Immediately discontinue use and quarantined any stock with the product/lot number described above in this recall notification. ii. If you have distributed any of these products, please immediately contact your customers, advise them of the recall situation and have them return their outstanding recalled stock to you. iii. Contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization (RGA) Number. Hours of operation are Monday through Friday, 8:00 a.m. to 4:30 p.m. EST. i. Once you have received the RGA Number, please enter it in the space provided on the Customer Product Recall Acknowledgement Form: ii. Return the completed form via email to Intersurgical Inc. at [email protected]. iii. The subject line should contain your Company name and RGA number. 1. This will allow us to document your receipt of this letter and the amount of product you have on-hand for return. iv. Intersurgical will credit your account when the product is received. v. Please note: If you do not have the recalled product in stock, please check the appropriate box on the Customer Product Recall Acknowledgement Form. vi. Return the completed form via email to Intersurgical Inc. at [email protected]. vii. The subject line should contain your Company name and Intersurgical Recall. Contact :315-451-2900 ext. 500 or email: [email protected].

Recall details

Recall number
Z-0096-2023
Status
Terminated
Initiated by the company
Aug 29, 2022
Posted by the FDA
Oct 17, 2022
Recalling firm
Intersurgical Inc
Quantity
340 pieces
Product code
CAE
Distribution
OH, TN, TX

Product description

Intersurgical Solus Standard, laryngeal mask airway, size 3, small adult, 30-50kg Model Number: 8003000

Code information (models, lots, serial numbers)

UDI-DISingle unit: 5030267073511 UDI-DICase: 05030267046959 Lot code: 32204071

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.