VENCLOSE, VENCLOSE PROCEDURE PACK, 7VM, REF VC-PPH-67A.… recallZ-0088-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated Sep 1, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Impacted Venclose Procedure Packs contain a 12cm introducer/sheath component, either component #128626 - SET INTRO 7FX12CM FITS 0.018 or component #137339 - SET INTRO 6.5FX12CM FITS 0.018, instead of a 7cm introducer/sheath, component #137340 - SET INTRO 6.5FX7CM FITS 0.018

FDA root cause category: Under Investigation by firm

What the company is doing

Medline Industries, LP sent a MEDICAL DEVICE RECALL notice to its consignees on 09/08/2022 by US Mail. The notice identified the product, problem, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers listedon the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list thequantity of affected product you have in inventory on the form. Even if you do not have anyaffected product in inventory, please complete and submit the form. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-22-131-FG Recall Code: RXXXXXXX 3.Upon receipt of your submitted response form, return labels and RGA information will beprovided, if applicable. Your account will receive credit when the returned product is received. 4.If you are a distributor, or have resold or transferred this product to another company orindividual, you are required by FDA regulations to notify them of this recall communication.Have the customers return any affected product to you for collection and return to MedlineIndustries, LP. If you have any questions, please contact the Recall Department at 866-359-1704."

Recall details

Recall number
Z-0088-2023
Status
Terminated
Initiated by the company
Sep 1, 2022
Posted by the FDA
Oct 13, 2022
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
1760 kits
Product code
OFF
Distribution
CA

Product description

VENCLOSE, VENCLOSE PROCEDURE PACK, 7VM, REF VC-PPH-67A. Cardiovascular procedure convenience kit.

Code information (models, lots, serial numbers)

UDI/DI 40193489423090, lot number 22EBV345, exp. 08/31/2024; UDI/DI 10193489423099, lot number 22FBF358, exp. 08/31/2024

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.