Skip to content

Atellica CH Lithium_2 (LITH_2), Siemens Material Number… recallZ-0104-2023 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 19, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting treatment efficacy.

FDA root cause category: Under Investigation by firm

What the company is doing

On September 19, 2022, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers should take the following actions: -Please review this letter with your Medical Director. -Discontinue use of and discard the affected product. Siemens is temporarily discontinuing production and distribution of the Atellica CH lithium reagents until the issue is resolved. In the interim, Siemens is recommending that customers transition to an alternate lithium assay.

Recall details

Recall number
Z-0104-2023
Status
Open, Classified
Initiated by the company
Sep 19, 2022
Posted by the FDA
Oct 20, 2022
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1342 US; 2266 OUS
Product code
NDW
FDA submissions
K050374
Distribution
Worldwide distribution - US Nationwide.

Product description

Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay

Code information (models, lots, serial numbers)

UDI-DI: 00630414287935 All lots impacted: Lots 110238 exp 10/18/22 110302 exp 12/13/22 120017 exp 4/18/23

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K050374 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90900).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page