ARTIS pheno, Interventional fluoroscopic x-ray system… recallZ-0119-2023 · Siemens Medical Solutions USA, Inc · initiated Sep 20, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens has become aware of three potential software issues with ARTIS pheno systems with software version VE10B. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be continued on an alternate system.
FDA root cause category: Software design
What the company is doing
On September 20, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. This notification covered three issues. 1. System Error Management, in which the operator is unable to release X-Ray in "Bypass fluoroscopy" mode which remains permanently inhibited, and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system. Customers should ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. 2. Head Holder Symbols, in which the user cannot tell whether a head holder has been selected in the system, may lead to a collision between the C-Arm and the head holder during system movement if the head holder is being used but not selected on the system. Customers should ensure that the configuration of the head holder is correct prior to system movements and double check configuration in the control room in case of uncertainty during the procedure. 3. Stand Movement, in which rarely no stand movement is possible after a routine brake test, leading to a possible delay of procedure. Customers should carry out routine brake tests with sufficient time before starting any procedure. Siemens will correct the software errors and contact customers to arrange a date to perform the corrective action. Please feel free to contact Siemens' service organization for an earlier appointment at 1-800-888-7436.
Recall details
- Recall number
- Z-0119-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 20, 2022
- Posted by the FDA
- Oct 21, 2022
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 140 US
- Product code
- JAA
- FDA submissions
- K163286
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.
Product description
ARTIS pheno, Interventional fluoroscopic x-ray system, Model No. 10849000
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI-DI: 04056869046877 Serial No 164140 164138 164159 164331 164332 164033 164287 164282 164210 164183 164184 164311 164312 164328 164034 164391 164318 164244 164306 164155 164340 164189 164271 164277 164173 164292 164231 164095 164057 164054 164055 164132 164329 164072 164190 164048 164025 164026 164027 164338 164094 164241 164326 164267 164186 164068 164121 164157 164019 164152 164161 164013 164069 164213 164073 164257 164293 164263 164298 164060 164083 164113 164021 164066 164362 164398 164124 164125 164366 164279 164075 164074 164084 164086 164098 164153 164330 164389 164394 164172 164295 164353 164234 164321 164242 164270 164275 164315 164344 164163 164158 164053 164171 164297 164134 164253 164333 164188 164377 164063 164064 164397 164299 164300 164105 164160 164024 164080 164049 164240 164359 164227 164228 164018 164022 164250 164061 164187 164115 164399 164011 164204 164205 164211 164212 164216 164100 164097 164269 164303 164322 164400 164196 164355 164352 164168 164319 164201 164202 164371
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS pheno 10849000Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system14 recalls64 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page