GORE CARDIOFORM ASD Occluder recallZ-0136-2023 · W L Gore & Associates, Inc. · initiated Sep 21, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from W L Gore & Associates, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing records (Release Test Results) indicating "Failed".
FDA root cause category: Employee error
What the company is doing
Between 09/21-32/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL/FIELD SAFETY NOTICE" via FedEx 2nd Day (U.S.), Postal Mail (EMEA), and Personal Delivery (Japan), informing customers that W.L. Gore & Associates (Gore) identified devices that were incorrectly released following a documented failure of a subset of quality tests. A failure of the subset of quality tests could potentially result in increased risks during the release of the device from the delivery system for the affected devices. Customer are instructed to: Identify and return any unused devices within the scope of this recall. Complete and sign the enclosed CUSTOMER RESPONSE FORM and return to [email protected] within 2 weeks of receipt of this notification. Share this letter with others in their institution as appropriate. Transfer this notice to other organization(s) as appropriate. If a listed device has been used, there is no patient follow-up needed and no further actions required other than informing Gore the device was used. Indicate the used device(s) on the CUSTOMER RESPONSE FORM and return to [email protected]. For questions, contact Gore Customer Service via email - [email protected]
Recall details
- Recall number
- Z-0136-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 21, 2022
- Posted by the FDA
- Oct 26, 2022
- Recalling firm
- W L Gore & Associates, Inc.
- Quantity
- 98 units
- Product code
- MLV
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GA, IA, ID, IL, ME, MN, NC, NY, JPA, TN, TX, WA, and WI. The countries of Denmark, and Japan.
Product description
GORE CARDIOFORM ASD Occluder. cardiovascular implant.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Numbers: ASD37A UDI-DI Codes: 00733132636501 Serial Numbers: 24468090 24468092 24468093 24468096 24468097 24468098 24468099 24468103 24468104 24480314 24480315 24480316 24480318 24468102 24468094 24468100 24468101 24480317 24468091 24468095 Model Number: ASD44A UDI-DI Code: 00733132636518 Serial Numbers: 24281498 24281499 24281500 24281501 24281502 24281507 24281509 24281511 24281512 24429270 24429271 24429276 24429278 24429279 24480351 24480352 24480353 24545359 24550148 24550149 24578446 24578447 24281506 24429272 24429273 24429274 24429277 24545360 24545362 24550147 24550150 24550151 24578443 24578445 24281503 24281504 24281505 24281508 24281510 24429275 24480354 24480355 24545358 Model Number: ASD44E UDI-DI Code: 00733132636617 Serial Numbers: 24305083 24305081 24550693 24550694 24305079 24305080 24305082 24550695 24550696 24550697
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- GORE CARDIOFORM ASD Occluder ASD37AW. L. Gore & Associates, Inc. · Transcatheter Septal Occluder1 recall42 reports
- GORE CARDIOFORM ASD Occluder ASD44AW. L. Gore & Associates, Inc. · Transcatheter Septal Occluder1 recall24 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page