Skip to content

Zio AT Clinical Reference Manual, Part: ALB0031, and Zio AT… recallZ-0183-2023 · iRhythm Technologies, Inc. · initiated Sep 28, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from iRhythm Technologies, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ambulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete; 2) Patches limited to 100 symptomatic and 500 asymptomatic wireless transmissions; once the maximum transmission limit is reached for either type, any further transmissions for that type will cease. Risk of delayed reporting to health providers.

FDA root cause category: Device Design

What the company is doing

On 9/28/2022, Field Safety Notice Labeling Corrections were sent to customers who were informed of labeling changes and that the firm intends to enhance the process of notifying patients that they are approaching the maximum limit for either symptomatic or asymptomatic transmissions by auto shipping another Zio AT patch to the patient. Customers are asked to do the following: 1) Please review this letter and disseminate it to the appropriate system providers and any personnel at your facility. Please share this notification with all device users within your facility to ensure they are aware of this notice. 2) Complete and return the customer reply form. Further questions can be directed to customer care at (888) 693-2401 or [email protected] or your local sales representative.

Recall details

Recall number
Z-0183-2023
Status
Open, Classified
Initiated by the company
Sep 28, 2022
Posted by the FDA
Nov 4, 2022
Recalling firm
iRhythm Technologies, Inc.
Quantity
1,443
Product code
DSI
FDA submissions
K163512
Distribution
US: MD, NY, FL, TN, WA, AL, MA, AZ, CA, KS, HI, PA, NJ, TX, MN, VA, AR, OR, NC, CO, KY, MS, LA, MI, WI, WV, IN, ID, SC, CT, SD, IA, NV, OK, OH, IL, VT, NH, NE, NM, MO, ME, DC, RI, MT, DE, GA, ND

Product description

Zio AT Clinical Reference Manual, Part: ALB0031, and Zio AT Important Information Pamphlet, Part: ALB0034. For Zio Ambulatory ECG monitoring system

Code information (models, lots, serial numbers)

UDI-DI: 00869770000210; Model/Revision: ALB0031/07, ALB0034/03

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page