Skip to content

Visualase recallZ-0244-2023 · Medtronic Navigation, Inc. · initiated Oct 14, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Navigation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a defect in the outer pouch sterile seal

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 10/03/2022, the firm verbally notified affected customers and on 10/14/2022 sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customer informing them that Medtronic identified a breach to the sterile barrier of the outer pouch of specific lots of VCLAS 9735560 .4MM Core Fiber 10MM Tip products. Customer are instructed to: immediately identify and segregate affected products within your inventory. Medtronic will contact you to facilitate product replacement when unaffected product is available. Medtronic has determined that the inner pouches maintain sterility, accordingly, you may use segregated products ensuring the following sterile technique workflow is followed: 1. Open the affected outer pouch outside of the sterile field and remove the smaller internal pouches. 2. Per the Instruction for Use: Prior to use, examine the product packaging (in this instance the inner pouch) for damage, deterioration, and expired shelf life. If found, do not use the product. 3. Open the internal pouches using sterile technique and carefully transfer the sterile contents into the sterile field. For questions, contact Medtronic Technical Service at 1-888-826-5603 or email at [email protected]

Recall details

Recall number
Z-0244-2023
Status
Open, Classified
Initiated by the company
Oct 14, 2022
Posted by the FDA
Nov 17, 2022
Recalling firm
Medtronic Navigation, Inc.
Quantity
95 units
Product code
ONO
FDA submissions
K181859, K211269, K053087
Distribution
U.S. Nationwide distribution including in the states AZ, CA, CO, DC, FL, IL, KY, MA, MD, ME, MI, MN, MO, NH, NJ, NY, OH, PA, RI, TX, VA, WA, and WV.

Product description

Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necrotize or coagulate soft tissue by physicians in minimally invasive surgical procedures.

Code information (models, lots, serial numbers)

Product Number/CFN: 9735560 UDI-DI Code: 00763000416430 Lot Numbers: 0224382403, 0224779378, 0224824522, 0224824523, 0224824525, 0224824767, 0224824768, 0224824769

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K181859, K211269, K053087 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page