2008T and 2008TBluestar Hemodialysis Machines with… recallZ-0734-2023 · Fresenius Medical Care Holdings, Inc. · initiated Jul 27, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

FDA root cause category: Process design

What the company is doing

Fresenius issued Updated Urgent Medical Device Correction REV B Customer Notification letter (Field Action Number: FA-2020-09-W _REV B) to consignees that have CDX equipped machines on 07-27-2022 via USPS First Class Mail. Letter states reason for recall, health risk and action to take: First, a software issue in SW versions 2.72 and earlier. All machines should be updated to current software (SW version 2.74), pursuant to the July 20, 2021, letter. Second, hardware issue that affects only CDX machines. To reduce the chance of EMI causing a false Remove USB Device 2 alarm from occurring we suggest the following: " Insert a USB device in USB Port 2 that will continually draw current. For example, a blank USB memory stick or a USB powered card reader. o NOTE: DO NOT use passive USB devices in USB Port 2 that go into a suspended state drawing minimal/low current which can trip the sensing circuit. An example would be a USB corded optical mouse. Complete/Return the response form enclosed.

Recall details

Recall number
Z-0734-2023
Status
Open, Classified
Initiated by the company
Jul 27, 2022
Posted by the FDA
Dec 20, 2022
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205
Product code
KDI
FDA submissions
K173972
Distribution
US Nationwide distribution.

Product description

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

Code information (models, lots, serial numbers)

Software issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K173972 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.