Skip to content

2008 Series Hemodialysis MachinesRecalls & reported problems across all models

Fresenius Usa, Inc. · 10 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

13 open recalls name at least one 2008 Series Hemodialysis Machines model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,768
Deaths reported
76
Injuries reported
133
Recalls
15

We found 10 models of 2008 Series Hemodialysis Machines by Fresenius Usa, Inc. in the FDA's device registry (GUDID). Together they are named in 1,768 adverse event reports in the FDA's MAUDE database, received between Jul 27, 2015 and Sep 23, 2026 — classified as 76 deaths, 133 injuries, 1,559 malfunctions.

The most frequently reported specific device problem is “Thermal Decomposition of Device” (801 reports).

15 FDA recall records name at least one of these models by its device identifier, 13 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 15 reportsDec 2023: 6 reportsJan 2024: 11 reportsFeb 2024: 6 reportsMar 2024: 19 reportsApr 2024: 11 reportsMay 2024: 10 reportsJun 2024: 4 reportsJul 2024: 7 reportsAug 2024: 10 reportsSep 2024: 10 reportsOct 2024: 13 reportsNov 2024: 12 reportsDec 2024: 5 reportsJan 2025: 12 reportsFeb 2025: 10 reportsMar 2025: 5 reportsApr 2025: 7 reportsMay 2025: 6 reportsJun 2025: 8 reportsJul 2025: 8 reportsAug 2025: 9 reportsSep 2025: 8 reportsOct 2025: 6 reportsNov 2025: 4 reportsDec 2025: 4 reportsJan 2026: 4 reportsFeb 2026: 4 reportsMar 2026: 3 reportsApr 2026: 9 reportsMay 2026: 4 reportsJun 2026: 5 reportsJul 2026: 4 reportsAug 2026: 7 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

2008 Series Hemodialysis Machines recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 8, 2024Class IIOpen, Classified

    Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.

    Z-3124-2024 · Root cause: Under Investigation by firm · names 5 of these models · Recall details · FDA record ↗

  2. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2347-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  3. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2346-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  4. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2345-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  5. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2344-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  6. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2343-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  7. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2341-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  8. Jun 3, 2024Class IIOpen, Classified

    An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.

    Z-2340-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

  9. Sep 6, 2023Class IOpen, Classified

    Potential PCBA leaching from tubing of hemodialysis machines.

    Z-0031-2024 · Root cause: Device Design · Recall details · FDA record ↗

  10. Sep 6, 2023Class IOpen, Classified

    Potential PCBA leaching from tubing of hemodialysis machines.

    Z-0030-2024 · Root cause: Device Design · Recall details · FDA record ↗

  11. Sep 6, 2023Class IOpen, Classified

    Potential PCBA leaching from tubing of hemodialysis machines.

    Z-0029-2024 · Root cause: Device Design · Recall details · FDA record ↗

  12. Sep 6, 2023Class IOpen, Classified

    Potential PCBA leaching from tubing of hemodialysis machines.

    Z-0028-2024 · Root cause: Device Design · Recall details · FDA record ↗

  13. Jul 27, 2022Class IIOpen, Classified

    (1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

    Z-0734-2023 · Root cause: Process design · names 2 of these models · Recall details · FDA record ↗

  14. Dec 14, 2017Class IITerminated

    When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.

    Z-1278-2018 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  15. Oct 5, 2017Class IITerminated

    While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

    Z-0829-2018 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗

Most reported 2008 Series Hemodialysis Machines problems

Device problems

  • Thermal Decomposition of Device801
  • Inadequate Ultra Filtration265
  • Fluid/Blood Leak257
  • Melted237
  • Adverse Event Without Identified Device or Use Problem161
  • Smoking91
  • Reflux within Device55
  • Mechanical Problem54

Patient problems

  • No Patient Involvement605
  • No Clinical Signs, Symptoms or Conditions475
  • No Known Impact Or Consequence To Patient339
  • Hemorrhage/Blood Loss/Bleeding109
  • Cardiac Arrest78
  • Death54
  • Loss of consciousness37
  • Low Blood Pressure/ Hypotension36

2008 Series Hemodialysis Machines models

10 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent 2008 Series Hemodialysis Machines adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Fresenius Usa, Inc. products

All Fresenius Usa, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page