6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use… recallZ-0917-2023 · The Anspach Effort, Inc. · initiated Dec 7, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from The Anspach Effort, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
FDA root cause category: Under Investigation by firm
What the company is doing
DePuy Synthes initiated "URGENT: MEDICAL DEVICE CORRECTION (Notification)" via UPS on 12/07/22. Letter states reason for recall, health risk an action to be taken. The customers were instructed to do the following: Review your inventory and make a plan to follow the suggested manufacturer inspection interval of 12 months associated with your product, as specified in the IFU. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. " Once this notification is received, please complete, sign, and return the attached Business Response Form (page 3 of this letter) via Fax: 877-496-5042 or Scan/email: [email protected] within 3 business days of receipt of this notification. Contact your Product Support Team through email at [email protected] or via phone at 1-800.327.6887, option #2 for additional support as needed.
Recall details
- Recall number
- Z-0917-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 7, 2022
- Posted by the FDA
- Jan 7, 2023
- Recalling firm
- The Anspach Effort, Inc.
- Quantity
- 6059 units in total
- Product code
- HBC
- FDA submissions
- K133604, K011444
- Distribution
- US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada
Product description
6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and cranium by trained medical/surgical personnel
Code information (models, lots, serial numbers)
GTIN: 00845384001737 All lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA CRANI-A-01Anspach Effort, The · MOTOR, DRILL, ELECTRIC1 recall0 reports
Other devices under the same FDA submission
46 devices cleared under K133604, K011444 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
- NA CRANI-A-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC1 recall358 reports
- NA QD8Synthes GmbH · MOTOR, DRILL, ELECTRIC386 reports
- NA SHORTSynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE343 reports
- NA QD11-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC312 reports
- NA QD8-G1Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE171 reports
- NA MEDIUMSynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE135 reports
- NA SHORT-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC134 reports
- NA LONGSynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE130 reports
- NA CRANI-LSynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall77 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91231).
- Z-0916-2023 · Open, Classified6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
- Z-0918-2023 · Open, ClassifiedAdult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
- Z-0919-2023 · Open, Classified6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
- Z-0920-2023 · Open, Classified7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
- Z-0921-2023 · Open, ClassifiedAdult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
- Z-0922-2023 · Open, Classified7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
- Z-0923-2023 · Open, Classified6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
- Z-0924-2023 · Open, ClassifiedPediatric Craniotome Ref:CRANI-P-G1
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page