Skip to content

Veritas Advanced Infusion Packs, REF: VRT-AI recallZ-1005-2023 · Johnson & Johnson Surgical Vision, Inc. · initiated Dec 7, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Johnson & Johnson Surgical Vision, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

FDA root cause category: Process control

What the company is doing

On 12/07/22, recall notices were sent to customers who were asked to do the following: 1) Discontinue using affected product. 2) Return product to the recalling firm. 3) Complete and return the Customer Reply form via email to [email protected] 4) Distributors, distribute the recall notice to your customers. Customers with questions or concerns can call the firm at 1-877-266-4543 Option 2.

Recall details

Recall number
Z-1005-2023
Status
Open, Classified
Initiated by the company
Dec 7, 2022
Posted by the FDA
Jan 20, 2023
Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Quantity
3133
Product code
HQC
FDA submissions
K203060
Distribution
US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

Product description

Veritas Advanced Infusion Packs, REF: VRT-AI

Code information (models, lots, serial numbers)

UDI-DI/Lots: 05050474700901/60400378, 60401981, 60401982

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 91290).