VERITAS ADVANCED INFUSION PACK VRT-AIRecalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · VERITAS ADVANCED INFUSION PACK · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- May 30, 2023Class IIOpen, Classified
Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Z-2247-2023 · Root cause: Process control · Recall details · FDA record ↗
- Dec 7, 2022Class IIOpen, Classified
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Z-1005-2023 · Root cause: Process control · Recall details · FDA record ↗
- Sep 1, 2022Class IIOpen, Classified
As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.
Z-0086-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
FDA clearance and classification
The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VERITAS ADVANCED INFUSION PACK VRT-AI
Has the VERITAS ADVANCED INFUSION PACK VRT-AI been recalled?
Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-2247-2023 (Class II), began on May 30, 2023: Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Who makes the VERITAS ADVANCED INFUSION PACK VRT-AI, and is it still sold?
The FDA's device database lists Johnson & Johnson Surgical Vision, Inc. as the company responsible for it (brand VERITAS ADVANCED INFUSION PACK, model VRT-AI). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation723 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation631 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation550 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls399 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls303 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation367 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page