No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 30, 2023Open, Classified

    Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

    Z-2247-2023 · Root cause: Process control · FDA record ↗

  2. Dec 7, 2022Open, Classified

    Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

    Z-1005-2023 · Root cause: Process control · FDA record ↗

  3. Sep 1, 2022Open, Classified

    As a result of returned product complaints for Veritas Phaco packs related to broken or cracked irrigation luer due to a manufacturing issue.

    Z-0086-2023 · Root cause: Under Investigation by firm · FDA record ↗