Skip to content

NuVasive MD Pulse III Multimodality System REF 1881500 recallZ-1052-2023 · NuVasive Inc · initiated Jul 19, 2022

Class III Open, Classified All recalls posted in February 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from NuVasive Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to an incorrect security key, customers were unable to connect to remote monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremity procedures.

FDA root cause category: Software design

What the company is doing

Between 07/19/2022 to 12/02/2022, affected customers were contacted and verbally informed during a software update that there was an issue related to the inability to remotely connect their system and does not allow physicians who is not physically in the operating room to be able to review live neuromonitoring data remotely and provide analysis to the surgeon in the operating room. The firm is planning on sending an "URGENT: MEDICAL DEVICE RECALL" Letter on 01/13/2023, to officially inform customers that a recall is being conducted due to a remote monitoring issue that may potentially not allow a physician who is not physically in the operating room to be able to review live neuromonitoring data remotely and provide analysis to the surgeon in the operating room. Customer Instructions: -Nuvasive, Inc. will contact a Representative at the hospitals to schedule the software key upgrade. -If the security key has been updated, no further action is necessary. For questions or assistance, contact Sr. Director, Global Quality at 858-336-3421 or email [email protected]

Recall details

Recall number
Z-1052-2023
Recall class
Class III
Status
Open, Classified
Initiated by the company
Jul 19, 2022
Posted by the FDA
Feb 3, 2023
Recalling firm
NuVasive Inc
Quantity
7
Product code
OLO
FDA submissions
K210574
Distribution
U.S. Nationwide distribution in the states of CA, FL, NC, OH, and TX.

Product description

NuVasive MD Pulse III Multimodality System REF 1881500 Pulse Software version 369-BL.US and 369-BLN.US

Code information (models, lots, serial numbers)

Part Number: 1881500 UDI Code: 0887517PLS137YJ Serial Numbers: PM00017 SM90018 SM90013 SM90021 PM00007 SM90032 SM90030

Other devices under the same FDA submission

1 device cleared under K210574 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page