Medical device recalls: February 2023
The FDA posted 140 medical device recall records in February 2023, from 40 recall events by 31 companies. 12 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- MEDLINE INDUSTRIES, LP - Northfield · 4 events, 64 records
- Access Vascular, Inc · 3 events, 5 records
- Covidien, LP · 2 events, 12 records
- Siemens Healthcare Diagnostics, Inc. · 2 events, 7 records
- Bard Peripheral Vascular Inc · 2 events, 2 records
- Olympus Corporation of the Americas · 2 events, 2 records
- Ge Medical Systems Israel, Functional Imaging · 1 event, 6 records, 6 Class I
- Siemens Medical Solutions USA, Inc · 1 event, 5 records
FDA root cause
- Device Design · 46
- Under Investigation by firm · 40
- Process control · 15
- Packaging process control · 11
- Nonconforming Material/Component · 8
- Software design · 6
- Employee error · 4
- Incorrect or no expiration date · 3
Class I recalls (12)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1137-2023 · Draeger Medical, Inc. · posted Feb 24, 2023 · Open, Classified
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
- Z-1138-2023 · Draeger Medical, Inc. · posted Feb 24, 2023 · Open, Classified
The magnets used in the swivel frame/tube set and the headgear clips might impact medical devices (implanted as well as not implanted) and metallic implants.
- Z-1139-2023 · Baxter Healthcare Corporation · posted Feb 24, 2023 · Open, Classified
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
- Z-1140-2023 · Baxter Healthcare Corporation · posted Feb 24, 2023 · Open, Classified
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
- Z-1141-2023 · Baxter Healthcare Corporation · posted Feb 24, 2023 · Open, Classified
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
- Z-1142-2023 · Baxter Healthcare Corporation · posted Feb 24, 2023 · Open, Classified
There is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
- Z-1053-2023 · Ge Medical Systems Israel, Functional Imaging · posted Feb 8, 2023 · Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
- Z-1054-2023 · Ge Medical Systems Israel, Functional Imaging · posted Feb 8, 2023 · Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
- Z-1055-2023 · Ge Medical Systems Israel, Functional Imaging · posted Feb 8, 2023 · Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
- Z-1056-2023 · Ge Medical Systems Israel, Functional Imaging · posted Feb 8, 2023 · Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
- Z-1057-2023 · Ge Medical Systems Israel, Functional Imaging · posted Feb 8, 2023 · Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
- Z-1058-2023 · Ge Medical Systems Israel, Functional Imaging · posted Feb 8, 2023 · Open, Classified
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
Class II recalls (125)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- BEQ-HLS 5050 USAClass IIZ-1188-2023 · Maquet Medical Systems USA · posted Feb 28, 2023 · Open, Classified
Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.
- Z-1183-2023 · Covidien, LP · posted Feb 27, 2023 · Open, Classified
Collagen film was placed on the opposite side of the mesh may cause delay to treatment/therapy, adhesions, erosion/migration, pain, fistula, hernia (recurrence), and failure of implant
- Z-1184-2023 · Stryker Corporation · posted Feb 27, 2023 · Open, Classified
Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage…
- Z-1185-2023 · Stryker Corporation · posted Feb 27, 2023 · Open, Classified
Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage…
- Triton Canister SoftwareClass IIZ-1186-2023 · Stryker Corporation · posted Feb 27, 2023 · Open, Classified
Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage…
- Z-1187-2023 · Encore Medical, LP · posted Feb 27, 2023 · Open, Classified
Tibial insert implant large boxes were labeled small and the small boxes were labeled large. If the incorrect labeling is identified this could lead to surgical delay and if the surgeon does not recognize the size discrepancy between the insert label and…
- Z-1172-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Monofilament Nylon Suture (McKesson Branded) and Coated Braided Silk Suture (McKesson Branded)Class IIZ-1173-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1174-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1175-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1176-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1177-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Surgilon Braided Nylon sutureClass IIZ-1178-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1179-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1180-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1181-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Custom Suture Pack ↗Class IIZ-1182-2023 · Covidien, LP · posted Feb 25, 2023 · Open, Classified
Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact
- Z-1169-2023 · Wright Medical Technology, Inc. · posted Feb 24, 2023 · Open, Classified
Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425,…
- Z-1170-2023 · Wright Medical Technology, Inc. · posted Feb 24, 2023 · Open, Classified
Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425,…
- Z-1171-2023 · Summit Medical, LLC · posted Feb 24, 2023 · Terminated
Sterile product pouches were not sealed.
- Z-1161-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 23, 2023 · Open, Classified
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
- Z-1162-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 23, 2023 · Open, Classified
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
- Z-1163-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 23, 2023 · Open, Classified
The convenience trays had improper placement of some of the components that were in the nonsterile portion of the tray, thus, not receiving sterilization.
- Z-1164-2023 · Siemens Medical Solutions USA, Inc · posted Feb 23, 2023 · Open, Classified
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1165-2023 · Siemens Medical Solutions USA, Inc · posted Feb 23, 2023 · Open, Classified
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1166-2023 · Siemens Medical Solutions USA, Inc · posted Feb 23, 2023 · Open, Classified
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1167-2023 · Siemens Medical Solutions USA, Inc · posted Feb 23, 2023 · Open, Classified
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1168-2023 · Siemens Medical Solutions USA, Inc · posted Feb 23, 2023 · Open, Classified
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
- Z-1155-2023 · Access Vascular, Inc · posted Feb 22, 2023 · Open, Classified
Product Mislabeled on the outer bag and inner kit Tyvek header bag
- Z-1156-2023 · Access Vascular, Inc · posted Feb 22, 2023 · Open, Classified
Label with the incorrect component listed on the inner kit Tyvek header bag
- Z-1157-2023 · Cardinal Health 200, LLC · posted Feb 22, 2023 · Open, Classified
Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
- Z-1158-2023 · Cardinal Health 200, LLC · posted Feb 22, 2023 · Open, Classified
Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
- Z-1159-2023 · Cardinal Health 200, LLC · posted Feb 22, 2023 · Open, Classified
Presource Kits contains the ICU MicroClave Clear Connector due to a manufacturing defect there is a gap between the connectors top and bottom housings which may lead to fluid leaks and loss of fluid path integrity
- Z-1144-2023 · Hologic, Inc. · posted Feb 21, 2023 · Open, Classified
C-Arm unexpected movement may cause blunt trauma should the tube arm impinge upon an individual
- Z-1149-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 21, 2023 · Open, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Z-1150-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 21, 2023 · Open, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Z-1151-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 21, 2023 · Open, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Z-1152-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 21, 2023 · Open, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Z-1153-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 21, 2023 · Open, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Z-1154-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 21, 2023 · Open, Classified
Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Z-1143-2023 · Siemens Healthcare Diagnostics, Inc. · posted Feb 17, 2023 · Open, Classified
Anti-HBe2 (aHBe2) interference with the Hepatitis B e Antigen (HBeAg) Assay, leads to elevated or reactive Hepatitis B e Antigen (HBeAg) results on samples that are non-reactive when HBeAg testing follows aHBe2 testing, regardless of whether both tests are…
- Z-1132-2023 · Olympus Corporation of the Americas · posted Feb 13, 2023 · Open, Classified
A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination
- Z-1133-2023 · Conformis, Inc. · posted Feb 13, 2023 · Open, Classified
Loctite thread locker (epoxy) may not have been applied to the screw that is part of the handle. In the absence of epoxy, the screw may loosen and fall into the patient during a procedure.
- Z-1134-2023 · Siemens Medical Solutions USA, Inc. · posted Feb 13, 2023 · Open, Classified
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
- Z-1135-2023 · Siemens Medical Solutions USA, Inc. · posted Feb 13, 2023 · Open, Classified
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
- Z-1136-2023 · Siemens Medical Solutions USA, Inc. · posted Feb 13, 2023 · Open, Classified
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
- Z-1109-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1110-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Surgical convenience kits labeled as: MEDLINE PERF. ASSOC. CUT DOWN TRAY, Reorder Number BM17BADClass IIZ-1111-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1112-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1113-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1114-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1115-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1116-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1117-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1118-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1119-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1120-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1121-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1122-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1123-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1124-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1125-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1126-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1127-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1128-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 10, 2023 · Open, Classified
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
- Z-1129-2023 · Access Vascular, Inc · posted Feb 10, 2023 · Open, Classified
Select catheter lots were identified to be labeled with incorrect expiration date, where the catheters expire prior to the labeled expiration date of the kit.
- Z-1130-2023 · Access Vascular, Inc · posted Feb 10, 2023 · Open, Classified
Select catheter lots were identified to be labeled with incorrect expiration date, where the catheters expire prior to the labeled expiration date of the kit.
- Z-1131-2023 · Access Vascular, Inc · posted Feb 10, 2023 · Open, Classified
Select catheter lots were identified to be labeled with incorrect expiration date, where the catheters expire prior to the labeled expiration date of the kit.
- Z-1107-2023 · Trackx Technology Inc · posted Feb 9, 2023 · Terminated
The device is breaking and falling off the Nuvasive T-Handle with Hudson Adapter on certain Nuvasive XLIF instruments during surgery when malleted.
- Z-1108-2023 · Bard Peripheral Vascular Inc · posted Feb 9, 2023 · Open, Classified
An increase in the reported complaint rate for inability to deploy was identified and related to poor slide block bonding.
- Z-1100-2023 · Quest Medical, Inc. · posted Feb 7, 2023 · Open, Classified
Delivery Sets may have tubing that could become detached that could result in patient infection.
- Z-1101-2023 · TriMed Inc. · posted Feb 7, 2023 · Open, Classified
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
- WishBone Medical Smart Correction Dual Joint Express Struts, external skeleton fixation device ↗Class IIZ-1102-2023 · Wishbone Medical, Inc. · posted Feb 7, 2023 · Open, Classified
Firm identified that user may be unable to adjust the length of the Smart Correction Dual Joint Express Struts due to the collar release pin backing out. This may lead to a delay in correction.
- SPiN Thoracic Navigation SystemClass IIZ-1103-2023 · Olympus Corporation of the Americas · posted Feb 7, 2023 · Open, Classified
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged procedure/surgery due to additional time to…
- Z-1104-2023 · Roche Molecular Systems, Inc. · posted Feb 7, 2023 · Open, Classified
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for…
- Z-1105-2023 · Roche Molecular Systems, Inc. · posted Feb 7, 2023 · Open, Classified
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for…
- Z-1106-2023 · Roche Molecular Systems, Inc. · posted Feb 7, 2023 · Open, Classified
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for…
- Z-1060-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1061-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1062-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1063-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1064-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1065-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1066-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1067-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1068-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1069-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1070-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1071-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1072-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1073-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1074-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1075-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1076-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1077-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1078-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1079-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1080-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1081-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1082-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1083-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1084-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1085-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1086-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1087-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1088-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1089-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1090-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1091-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1092-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1093-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1094-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1095-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1096-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1097-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 6, 2023 · Open, Classified
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
- Z-1098-2023 · Zimmer, Inc. · posted Feb 6, 2023 · Open, Classified
Two planers were found to be unusable during sawbones shoulder training.
- Z-1099-2023 · Teknimed Sas · posted Feb 6, 2023 · Open, Classified
Incorrect expiry date. Product is labeled with expiry date 2027-02-28. The correct expiration date is 2025-02-28.
- Z-1045-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 1, 2023 · Open, Classified
The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.
- Z-1046-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 1, 2023 · Open, Classified
The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.
- Z-1047-2023 · MEDLINE INDUSTRIES, LP - Northfield · posted Feb 1, 2023 · Open, Classified
The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.
- Z-1048-2023 · W L Gore & Associates, Inc. · posted Feb 1, 2023 · Open, Classified
Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder…
- Z-1049-2023 · NovaSignal Corp. · posted Feb 1, 2023 · Open, Classified
Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
- Z-1050-2023 · NovaSignal Corp. · posted Feb 1, 2023 · Open, Classified
Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
- Z-1051-2023 · Bard Peripheral Vascular Inc · posted Feb 1, 2023 · Open, Classified
Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).
Class III recalls (3)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1160-2023 · S & Mohr America, S.A. De C.V. · posted Feb 22, 2023 · Open, Classified
Incorrect expiration date on product label. The manufacturing date was listed as the expiration date. Correct expiration date is "22.07.2027"
- Z-1052-2023 · NuVasive Inc · posted Feb 3, 2023 · Open, Classified
Due to an incorrect security key, customers were unable to connect to remote monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremity procedures.
- Z-1059-2023 · Owens & Minor Distribution, Inc. · posted Feb 3, 2023 · Open, Classified
Face masks were placed into the incorrect dispenser boxes. The face masks come with ear loops, but the dispenser is incorrectly labeled as face masks with ties instead of ear loops.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.