Philips Garbin Ventilator, Model Number 1058180B recallZ-0884-2023 · Philips Respironics, Inc. · initiated Dec 7, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Respironics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.
FDA root cause category: Process design
What the company is doing
Telephone calls and URGENT Medical Device Recall notification letters were distributed beginning 12/6/22. Actions you must take as a device owner or healthcare provider to prevent risks for patients. Device is on patients: For affected Trilogy 100 and Trilogy 200 devices that are on patients, ensure that your clinical assessment of the patients is up to date, reflecting the true acuity of the patients and their respiratory therapy requirements. Your patient may experience serious harm if their device malfunctions and immediate intervention is not available. Until Philips Respironics is able to correct this problem you must adhere to the instructions below. Further instructions are provided below for devices that are not on patients. For ventilator dependent patients: 1. Consider transitioning ventilator dependent patients to an alternative life-support ventilator if, as determined by the clinical assessment for the individual patient in consultation with the patient s healthcare provider, a malfunction would lead to serious injury if timely access to a backup ventilator is not possible. 2. Quarantine all impacted devices that are removed from service. 3. Please indicate which Serial Numbers must be removed from service in exchange for an alternative device For all other patients: 1. Ensure that device alarm settings are configured appropriately for your patient s needs, with special consideration for the alarms listed below. Ensure that alternative ventilation equipment is available, as appropriate to patient need, in case of emergency issues with alarms that cannot be resolved. 2. Ensure that backup monitoring is configured as required by the clinical assessment. Pulse oximetry, ECG, heart rate, blood pressure, or respiratory rate may provide additional advanced warning of and lead to faster intervention if hypoxemia were to occur. If these accessories are prescribed, ensure patients advised of proper use. 3. Consider using an inline bacterial filter which ma
Recall details
- Recall number
- Z-0884-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 7, 2022
- Posted by the FDA
- Jan 6, 2023
- Recalling firm
- Philips Respironics, Inc.
- Quantity
- 9
- Product code
- CBK
- FDA submissions
- K083526
- Distribution
- US Nationwide and Global Distribution to countries of: Australia, Brazil, Canada, China, Germany, Japan, Korea, and Latin America.
Product description
Philips Garbin Ventilator, Model Number 1058180B
Code information (models, lots, serial numbers)
UDI: 00606959429338; Serial Numbers: GV119100314 GV118011626 GV117030343 GV117020301 GV11910210D GV117060237 GV11807062D GV119070802 GV11605061E
Other devices under the same FDA submission
23 devices cleared under K083526 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy 00Respironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE15 deaths2,059 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Trilogy Trilogy 100 Ventilator, ChinaRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 death447 reports
- Trilogy100 Ventilator, Japan-RentalRespironics, Inc. · Medical device data system2 deaths334 reports
- Trilogy Trilogy100 Ventilator, Latin AmericaRespironics, Inc. · Oximeter154 reports
- Trilogy 100 Trilogy100 Ventilator, Bluetooth, ItalyRespironics, Inc. · Medical device data system146 reports
- Trilogy 00Respironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE1 death135 reports
- Trilogy Trilogy 100 Ventilator, BrazilRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE2 deaths127 reports
- Trilogy 00Respironics, Inc. · Oximeter4 deaths119 reports
- Trilogy Trilogy100 Ventilator, AustraliaRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE97 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91293).
- Z-0882-2023 · Open, ClassifiedPhilips Trilogy 100 Ventilator, Model Numbers 1054260, 1054096, U1054260, U1054260B, 1054260B, CA1054096B, BT1054260, 1054096B, R1054260, RSL1054260, CA1054096, R1054260B, R1054655…
- Z-0883-2023 · Open, ClassifiedPhilips Trilogy 200 Ventilator, Model Numbers 1040005, U1040005, 1040005B, U1040005B, CA1032800B, CA1032800, KO1032800, 1032800, R1032804TP, R1032804B, R1032804TPV, 1032804B
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.