JACKSON-PRATT recallZ-0974-2023 · Cardinal Health 200, LLC · initiated Dec 14, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
Cardinal Health issued an URGENT MEDICAL DEVICE PRODUCT RECALL notice to its consignees on 12/14/2022 by overnight mail. The notice explained the issue, the risk, and requested the following: Actions Required: 1. REVIEW your inventory for the affected product codes and lots. Location of product codes and lots are shown above. 2. SEGREGATE and quarantine all affected product upon review of your inventory. Product should not be used and cease using the product immediately. 3. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this product recall notice and share a copy of this notice. 4. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether you have affected product. Return of Product and Available Assistance: CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5pm EST: . Hospital 800-964-5227 . Federal Government 800-444-1166 . Distributor 800-635-6021 . All other customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
Recall details
- Recall number
- Z-0974-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 14, 2022
- Posted by the FDA
- Jan 18, 2023
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- 1768 units
- Product code
- GBX
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.
Product description
Cardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2211
Code information (models, lots, serial numbers)
UDI/DI 50885380182486(cs), 20885380182485(ea), Lot Numbers: N210278, N210368, N210423, N210460, N210468 , N210469, N210543, N220050, N220099, N220139, N220163
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- JACKSON-PRATT JP-2211Cardinal Health 200, LLC · CATHETER, IRRIGATION1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 91322).
- Z-0968-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2187
- Z-0969-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2188
- Z-0970-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2189
- Z-0971-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2190
- Z-0972-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2191
- Z-0973-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2210
- Z-0975-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2212
- Z-0976-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2213
- Z-0977-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2214
- Z-0978-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2215
- Z-0979-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 7 mm, REF JP-2216
- Z-0980-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 mm, REF JP-2217
- Z-0981-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2221
- Z-0982-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2223
- Z-0983-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2225
- Z-0984-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2226
- Z-0985-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2227
- Z-0986-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2228
- Z-0987-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2229
- Z-0988-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2230
- Z-0989-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2231
- Z-0990-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2232
- Z-0991-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 15 Fr, REF JP-2233
- Z-0992-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 24 Fr, REF JP-2234
- Z-0993-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 28 Fr, REF JP-2290
- Z-0994-2023 · Open, ClassifiedCardinal Health Jackson-Pratt Channel Drain, 32 Fr, REF JP-2292
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page