Skip to content

Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF… recallZ-0989-2023 · Cardinal Health 200, LLC · initiated Dec 14, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

Cardinal Health issued an URGENT MEDICAL DEVICE PRODUCT RECALL notice to its consignees on 12/14/2022 by overnight mail. The notice explained the issue, the risk, and requested the following: Actions Required: 1. REVIEW your inventory for the affected product codes and lots. Location of product codes and lots are shown above. 2. SEGREGATE and quarantine all affected product upon review of your inventory. Product should not be used and cease using the product immediately. 3. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this product recall notice and share a copy of this notice. 4. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether you have affected product. Return of Product and Available Assistance: CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5pm EST: . Hospital 800-964-5227 . Federal Government 800-444-1166 . Distributor 800-635-6021 . All other customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.

Recall details

Recall number
Z-0989-2023
Status
Open, Classified
Initiated by the company
Dec 14, 2022
Posted by the FDA
Jan 18, 2023
Recalling firm
Cardinal Health 200, LLC
Quantity
8563 units
Product code
GBX
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.

Product description

Cardinal Health Jackson-Pratt Channel Drain, 19 Fr, REF JP-2231

Code information (models, lots, serial numbers)

UDI/DI 50885380182639(cs), 20885380182638(ea), Lot Numbers: N210278, N210347, N210368, N210423, N210460, N210468, N210469, N210543, N220050, N220099, N220139, N220163

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 91322).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page