Getinge Flow-e Anesthesia System, Part No. 6887900 recallZ-0961-2023 · Getinge Usa Sales Inc · initiated Dec 22, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
FDA root cause category: Software Design Change
What the company is doing
On December 22, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to examine their inventory immediately to determine if they have any of the affected serial numbers. If you have an affected system running on software version 4.8.0 through 4.8.3, you may continue use of the device until the software is updated with the following considerations: 1. If during use, the device is switched from Automatic to Manual and it is found that the manual bag is empty the user can activate the Emergency ventilation or turn off/on the device to reset the disabled valves function. 2. Users may also use the resuscitator that is available in the Operating Room. Getinge will update the software to correct the issue free of charge. You can arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9GETUSA / (888) 943 8872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Recall details
- Recall number
- Z-0961-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 22, 2022
- Posted by the FDA
- Jan 26, 2023
- Recalling firm
- Getinge Usa Sales Inc
- Quantity
- 8.016 Total (7.766 OUS; 250 US)
- Product code
- BSZ
- FDA submissions
- K191027
- Distribution
- US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.
Product description
Getinge Flow-e Anesthesia System, Part No. 6887900
Code information (models, lots, serial numbers)
Part # UDI 6887900 07325710010457 Serial Numbers 50012-50197 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
Other devices under the same FDA submission
5 devices cleared under K191027 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- Flow-i C20 Anesthesia System 6888520Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall2 deaths88 reports
- Flow-i C30 Anesthesia System 6888530Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall13 reports
- Flow-i C40 Anesthesia System 6888540Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall9 reports
- Vaporizer Sevoflurane, MAQUET filling 6886601Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91339).
- Z-0959-2023 · Open, ClassifiedGetinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400
- Z-0960-2023 · Open, ClassifiedGetinge Flow-c Anesthesia System, Part No. 6887700
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page