Flow-i C40 Anesthesia System 6888540Recalls & reported problems
Maquet Critical Care AB · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07325710010631), received between Mar 27, 2023 and Jul 1, 2025 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Output Problem” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
10 other recalls cover devices cleared under the same FDA submission (K191027); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 24, 2026Class IIOpen, Classified
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Z-2860-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Output Problem6
- Adverse Event Without Identified Device or Use Problem3
- Contamination1
- Medical Gas Supply Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Low Oxygen Saturation2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2025Contamination; Medical Gas Supply Problem; Contamination; Medical Gas Supply Problem
DURING THE INSPECTION OF THE ANESTHESIA WORKSTATION IT WAS DISCOVERED THAT THE AIR GAS MODULE WAS CONTAMINATED WITH WATER. THERE WAS NO PATIENT HARM. MANUFACTURER'S REF. #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22370198 ↗MalfunctionApr 30, 2025Adverse Event Without Identified Device or Use Problem; Output Problem; Adverse Event Without Identified Device or Use Problem; Output Problem
IT WAS REPORTED THAT WHEN USING THE LASER DURING THE END OF AN ENT SURGERY IN THE PATIENT'S EAR, A FLAME WAS GENERATED CAUSING A BURN IN THE PATIENT EAR. PATIENT FINAL OUTCOME WAS NO INJURY AND NO INTERVENTION REQUIRED. MANUFACTURER'S REFERENCE #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21933485 ↗MalfunctionOct 8, 2024Output Problem; Output Problem
MANUFACTURER'S REFERENCE #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20394707 ↗MalfunctionMay 23, 2024Output Problem; Output Problem
IT WAS REPORTED THAT THE SEVOFLURANE VAPORIZER IN THE ANESTHESIA WORKSTATION FAILED IN SYSTEM CHECKOUT AND GENERATED A TECHNICAL ERROR CODE INDICATING VAPORIZER ERROR. THE VAPORIZER IS THE UNIT CONTAINING ANESTHETIC AGENT IN THE ANESTHESIA WORKSTATION. WHEN THE VAPORIZER WAS RECEIVED FOR INVESTIGATION, A YELLOW/BROWN RESIDUE WAS FOUND INSIDE THE VAPORIZER. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REFERENCE #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19383116 ↗InjuryFeb 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT WHILE THE ANESTHESIA SYSTEM WAS IN USE FOR TREATMENT, THE PATIENT SATURATION DECREASED TO 77 %. THE PATIENT WAS DISCONNECTED FROM THE SYSTEM AND MANUALLY VENTILATED WITH AN AMBU BAG. THE PATIENT RECOVERED AND THE FINAL PATIENT OUTCOME WAS NO INJURY. MANUFACTURER'S REFERENCE #: (B)(4).
Patient problems: Low Oxygen Saturation
MAUDE report 18632120 ↗InjuryFeb 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT WHILE THE ANESTHESIA SYSTEM WAS IN USE FOR TREATMENT, THE PATIENT SATURATION DECREASED TO 77 %. THE PATIENT WAS DISCONNECTED FROM THE SYSTEM AND MANUALLY VENTILATED WITH AN AMBU BAG. THE PATIENT RECOVERED AND THE FINAL PATIENT OUTCOME WAS NO INJURY. MANUFACTURER'S REFERENCE #: (B)(4).
Patient problems: Low Oxygen Saturation
MAUDE report 18630705 ↗MalfunctionOct 17, 2023Output Problem; Output Problem
MANUFACTURER'S REFERENCE NUMBER: (B)(4). IT WAS REPORTED THAT THE ANESTHESIA SYSTEM DID NOT CYCLE PROPERLY. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17948284 ↗MalfunctionOct 2, 2023Output Problem; Output Problem
IT WAS REPORTED THAT THE ANESTHESIA SYSTEM FAILED THE LEAKAGE TEST DURING SYSTEM CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT. MANUFACTURER'S REFERENCE NUMBER: 886275
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17851510 ↗MalfunctionMar 27, 2023Output Problem; Output Problem
MANUFACTURE'S REFERENCE #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16620947 ↗
FDA clearance and classification
Cleared through 510(k) K191027 on Mar 18, 2020 (decision: Substantially Equivalent). Submitted by Maquet Critical Care AB as “Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System”. The FDA review took 335 days. FDA record ↗
The FDA regulates product code BSZ (gas-machine, anesthesia) under 21 CFR 868.5160 as a Class II device. 21 CFR 868.5160 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Recalls under the same FDA submission
These recalls involve devices cleared under K191027. One submission often covers many models, so check the product and lot details in each record.
- Z-1074-2025 · Jan 15, 2025 · Open, ClassifiedVaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
- Z-1877-2024 · May 1, 2024 · Open, ClassifiedVaporizer Sevoflurane Maquet Filling, Product code/REF number (6682282), Component of: Flow-i C20, Flow-i C30,Flow-i C40, Flow-c, Flow-e.
- Z-0960-2023 · Dec 22, 2022 · Open, ClassifiedGetinge Flow-c Anesthesia System, Part No. 6887700
- Z-0959-2023 · Dec 22, 2022 · Open, ClassifiedGetinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400
- Z-0961-2023 · Dec 22, 2022 · Open, ClassifiedGetinge Flow-e Anesthesia System, Part No. 6887900
- Z-0422-2023 · Nov 7, 2022 · Open, ClassifiedFlow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)
- Z-1279-2022 · May 2, 2022 · TerminatedFlow-e Anesthesia System : Intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887900
- Z-1278-2022 · May 2, 2022 · TerminatedFlow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887700
- Z-0468-2022 · Dec 9, 2021 · Open, ClassifiedVaporizer Sevoflurane Maquet Filling, Part No. 6886601 Component of the following systems: Product Code(s) UPN/UDI: Flow-c Anesthesia System 6887700 07325710009765 Flow-e Anesthe
- Z-1181-2021 · Jan 29, 2021 · TerminatedAnesthesia gas-machine - Product Usage: intended for use in administering anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in
Common questions about the Flow-i C40 Anesthesia System 6888540
Has the Flow-i C40 Anesthesia System 6888540 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2860-2026 (Class II), began on Jun 24, 2026: Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Flow-i C40 Anesthesia System 6888540?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Mar 27, 2023 and Jul 1, 2025. In those reports, the most common specific device problems are “Output Problem” (6), “Contamination” (1) and “Medical Gas Supply Problem” (1); the most common patient problem is “Low Oxygen Saturation” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Flow-i C40 Anesthesia System 6888540?
None of the reports are classified as deaths, 2 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Flow-i C40 Anesthesia System 6888540 FDA approved?
It was cleared through 510(k) K191027 on Mar 18, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Flow-i C40 Anesthesia System 6888540 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Flow-i C40 Anesthesia System 6888540, and is it still sold?
The FDA's device database lists Maquet Critical Care AB as the company responsible for it (brand Flow-i C40 Anesthesia System, model 6888540). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gas-Machine, Anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,073 reports
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,193 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,029 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls1,977 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,535 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls17 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page