Skip to content

GORE CARDIOFORM Septal Occluder recallZ-1048-2023 · W L Gore & Associates, Inc. · initiated Dec 14, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from W L Gore & Associates, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.

FDA root cause category: Employee error

What the company is doing

On 12/14/22, recall notices were mailed to customers informing them to take the following actions: 1) Identify and return any unused devices. 2) Complete and return the customer response form to to [email protected] 3) This notice needs to be shared with those who should be aware within your institution or to any organization where potentially affected devices have been transferred. Customers with questions can contact the firm at [email protected]

Recall details

Recall number
Z-1048-2023
Status
Open, Classified
Initiated by the company
Dec 14, 2022
Posted by the FDA
Feb 1, 2023
Recalling firm
W L Gore & Associates, Inc.
Quantity
14
Product code
MLV
Distribution
US Nationwide distribution in the states of NY, CA, FL, CT, TX, AR, IN, MA, MO, SD, VA.

Product description

GORE CARDIOFORM Septal Occluder, REF: GSX0030A

Code information (models, lots, serial numbers)

UDI-DI/Serial Numbers: 00733132631032/25556388, 25556389, 25556390, 25556391, 25556392, 25556393, 25556394, 25556398, 25556399, 25556400, 25556401, 25556402, 25556403, 25556406

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates