Skip to content

HydroMID 4Fr Single Lumen - Basic Kit, Model Number 80004002 recallZ-1129-2023 · Access Vascular, Inc · initiated Dec 16, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Access Vascular, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Select catheter lots were identified to be labeled with incorrect expiration date, where the catheters expire prior to the labeled expiration date of the kit.

FDA root cause category: Incorrect or no expiration date

What the company is doing

Urgent Medical Device Correction notification letters were sent to customers beginning 12/16/22.. Access Vascular has identified that the catheter packaged with specific lots of the HydroPICC Kits and HydroMID Kits have a shorter expiration date than what is listed on the outer kit package. These catheters are individually packaged and supplied sterile along with a sterile kit. Our records indicate that your facility has received kits from the affected lots. We are contacting you to provide information on action to take to ensure patient safety. - Product that has already been used: No action required. The catheter and all kit components are currently within their expiration date, so no patient safety risk has been identified. - Product that has not been used: The catheter and all kit components are currently, within their expiration date, so no immediate patient safety risk has been identified. However, please refrain from using these kits as using the product used past its expiration date may contaminate the sterile field and result in an infection. AVI will provide a replacement label with the correct expiration date or replace the affected products. The following page identifies the shipments you have received with these impacted lots. Please forward this notice in the event the affected product was further distributed. We will contact you to identify the number of kits that are currently in your inventory and determine the appropriate correction with your input, in a timely manner. Access Vascular is committed to providing the highest quality products and support and is implementing corrective actions to prevent this issue from recurring in the future. We apologize for any inconvenience this may cause you and appreciate your understanding as we take steps to correct this issue. If you have any questions about this communication, please contact us at 781-538-6594.

Recall details

Recall number
Z-1129-2023
Status
Open, Classified
Initiated by the company
Dec 16, 2022
Posted by the FDA
Feb 10, 2023
Recalling firm
Access Vascular, Inc
Quantity
1,789 HydroMID devices
Product code
FOZ
FDA submissions
K203069
Distribution
US Nationwide distribution in the states of CT, IL, IN, MA, NE, TN, and TX.

Product description

HydroMID 4Fr Single Lumen - Basic Kit, Model Number 80004002

Code information (models, lots, serial numbers)

UDI-DI: (01)00850030354006; Lot Number: 11428361, 11393260, 11395226

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K203069 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91426).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page