Skip to content

NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1.… recallZ-1050-2023 · NovaSignal Corp. · initiated Dec 16, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from NovaSignal Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 12/16/22, recall notices were emailed to customers. Customers were told they would be contacted by the recalling firm to facilitate the proper disposition of affected devices. If an unexpected reboot occurs, the recalling firm recommends pausing the use of affected devices until unaffected devices arrive. Complete and return the customer response form. Customers with questions can contact Customer Operations at [email protected]

Recall details

Recall number
Z-1050-2023
Status
Open, Classified
Initiated by the company
Dec 16, 2022
Posted by the FDA
Feb 1, 2023
Recalling firm
NovaSignal Corp.
Quantity
4
Product code
IYN
FDA submissions
K160442, K180455
Distribution
US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL

Product description

NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)

Code information (models, lots, serial numbers)

System (Tablet)UDI-DI/ Serial Numbers: 00850681007443/ 22040001, 00850681007443/ 22040002, 00850681007443/21021223 (00850681007221/ 21090003), 00850681007443/ 22064227 (00850681007221/ 21090004)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K160442, K180455 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 91467).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page