Skip to content

MEDLINE UNITE recallZ-1047-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated Dec 22, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.

FDA root cause category: Device Design

What the company is doing

Medline issued a MEDICAL DEVICE RECALL notice to its direct consignees on 01/05/2023 by US mail and its distributors on 01/07/2023 by email. The notice explained the problem and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected items. Quarantine all affected product. All consignment inventory will be addressed by your local Medline UNITE Sales Representative. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Even if you do not have any affected product please complete and return the form, as indicated on the response form. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP. FURTHER ACTION: Medline is immediately recommending the use of our 1.6mm pre-drill for 2.0mm screws, and our 1.9mm pre-drill for the use of our 2.4mm screws and 2.7mm screws. If you have any questions, please contact 866-359-1704."

Recall details

Recall number
Z-1047-2023
Status
Open, Classified
Initiated by the company
Dec 22, 2022
Posted by the FDA
Feb 1, 2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
190 units
Product code
KRR
FDA submissions
K181280
Distribution
US Nationwide distribution.

Product description

MEDLINE UNITE FOOT & ANKLE DRILL BIT, 1.3MM Solid Core, REF MPN10013

Code information (models, lots, serial numbers)

GTIN 10888277699113, Lot Number ?

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 91491).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page