Skip to content

Sprint 200 wheeled stretcher, Item numbers 1ES200-7… recallZ-1211-2023 · Linet Spol. S.r.o. · initiated Jan 18, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Linet Spol. S.r.o. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Labeling being updated to assure user is securing the siderail to a locked position to prevent potential patient fall and injury.

FDA root cause category: Labeling design

What the company is doing

The firm issued "URGENT MEDICAL DEVICE CORRECTION" recall letters dated 1/17/2023 via email on 1/18/2023. The letter explains the issue that an audible clicking sound can be heard when the side rail is raised, but this sound does not indicate the side rail has locked into the raised position. The consignee is to take immediate action to notify caregivers using the Sprint 200 of this potential hazard and proper use for securing a side rail in the raised position. The consignee is informed that Linet will be adding additional safety warning labels located at each side rail latch (2 per side of stretcher), add additional safety warning labels to both side rails (1 per side of stretcher), and that additional training material and an updated copy of the Sprint 200 Emergency Stretcher Instructions for Use and Technical Description manual is available using the QR code in the letter or by following a link to access the materials. The firm's service department will be contacting the consignees to schedule an appointment to perform the corrections. Photos of the additional warning labels are included in the letter. Please contact LINET service department at 877-815-9897 with immediate questions or concerns.

Recall details

Recall number
Z-1211-2023
Status
Open, Classified
Initiated by the company
Jan 18, 2023
Posted by the FDA
Mar 4, 2023
Recalling firm
Linet Spol. S.r.o.
Quantity
335 stretchers (310 US, 25 OUS)
Product code
FPO
Distribution
US Distribution was made to AL, CA, FL, IA, KY, LA, MN, NV, NC, OK, TX, VA, WA, and WV. There was no government/military distribution. Foreign distribution was made to Australia, Bahrain, Belgium, Germany, Great Britain, Italy, Netherland, Singapore, South Africa, Switzerland, and United Emirates Canada.

Product description

Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

1ES200-7 - Serial numbers 20220203852, 20220203853, 20220203854, 20220203407, 20220203429, 20220203425, 20220203427, 20220203428, 20220203430, 20220203836, 20220203864, 20220203402, 20220203404, 20220203420, 20220203422, 20220203841, 20220203850, 20220203858, 20220203862, 20220203865; UDI-DI 08592654358467. 1ES210-1 - Serial numbers 20220204642, 20220204649, 20220204654, 20220204666, 20220207624, 20220261655, 20220261670, 20220261671, 20220261672, 20220204637, 20220204638, 20220204639, 20220204641, 20220204646, 20220204651, 20220204656, 20220204657, 20220207626, 20220207630, 20220261659, 20220261660, 20220261650, 20220261661, 20220261662, 20220261663, 20220261664, 20220261665, 20220261666, 20220261667, 20220261668, 20220261669, 20220261676, 20220204665, 20220207622, 20220207631, 20220207634, 20220207637, 20220207638, 20220207641, 20220207642, 20220207644, 20220207645, 20220261648, 20220261649, 20220261651, 20220261656, 20220261673, 20220261674, 20220261677, 20220261678, 20220261679, 20220261680, 20220261681, 20220261682, 20220261683, 20220261657, 20220261675, 20220190550, 20220190552, 20220190554, 20220190548, 20220204648, 20220204662, 20220207621, 20220207623, 20220207628, 20220207632, 20220207636, 20220207643, 20220204643, 20220204647, 20220204650, 20220204652, 20220204653, 20220204655, 20220204658, 20220204659, 20220204661, 20220204663, 20220204664, 20220207616, 20220207617, 20220207618, 20220207619, 20220207620, 20220207627, 20220207629, 20220207633, 20220207639, 20220261654; UDI-DI 08592654356685. 1ES211-1 - Serial numbers 20220210341, 20220190528, 20220190539, 20220190540, 20220190544, 20220190530, 20220190537, 20220190543, 20220207647, 20220207648, 20220207649, 20220207650, 20220207651, 20220207653, 20220207655, 20220207659, 20220207661, 20220207662, 20220207663, 20220207664, 20220207666, 20220207667, 20220207668, 20220207670, 20220210334, 20220212559, 20220212547, 20220212548, 20220212553, 20220210306, 20220210308, 20220210315, 20220210317, 20220210318, 20220210320, 20220210330, 20220210336, 20220210337, 20220212545, 20220212558; UDI-DI 08592654356678.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.