V. Mueller recallZ-1236-2023 · Carefusion 2200 Inc · initiated Jan 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Carefusion 2200 Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product did not meet shelf-life testing requirements resulting in a breach of the sterility of the Genesis container holding surgical equipment.

FDA root cause category: Packaging change control

What the company is doing

The recalling firm issued letters to customers and a separate letter to distributors dated 1/30/2023 via email and FedEx. The customer letters described the problem, potential clinical impact, provided product identification including photographs, and provided actions the consignees are to take. Instructions included: (1) Discontinue use of the affected product. Check all inventory locations within their institution for the affected product and immediately destroy all remaining product in their possession. A list of the components and their material makeup are provided so the consignee can have them disposed of according to their local facility's process; (2) Share the notification with all users of the affected product within their facility network to ensure they are also aware of the recall; (3) If the product was purchased from a distributor, the consignee is to contact the distributor directly for further instructions and credit resolution; (4) Complete the attached Customer Response Form and return it to the BD contact noted on the form. The letter also informs the consignee they will no longer manufacture or distribute the Genesis Reusable Rigid Sterilization Containers (CD5 Series only). The distributor letters were similar except they were requested to identify their customers who purchased the affected product and provide a copy of the customer letter to them. The distributors were to return the Distributor Response Form to the recalling firm acknowledging receipt and that they have read and understand the recall notice.

Recall details

Recall number
Z-1236-2023
Status
Open, Classified
Initiated by the company
Jan 30, 2023
Posted by the FDA
Mar 9, 2023
Recalling firm
Carefusion 2200 Inc
Quantity
5,952 devices
Product code
KCT
FDA submissions
K142529
Distribution
Worldwide distribution - US Nationwide, including Puerto Rico. There was also government/military distribution. The countries of Australia, Canada, Denmark, Finland, Hong Kong, Ireland, Japan, Malaysia, Philippines, Poland, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom, United Arab Emirates.

Product description

CD5 Series of Genesis Reusable Rigid Sterilization Container Systems catalog number and product name: 1. CD5-3B, CONTAINER SHALLOW LG ENDO 3IN PERF; 2. CD5-3C, CONTAINER SHALLOW LG ENDO 3IN SOLID; 3. CD5-61B, GENESIS CONT LARGE ENDO PERF 28X11X6IN; 4. CD5-61C, GENESIS CONT LARGE ENDO SOLID 28X11X6IN; 5. CD5-3B-BLACK, GENESIS LG NARROW 3" PERF BOTTOM BLACK; 6. CD5-3B-BLUE, GENESIS LG NARROW 3" PERF BOTTOM BLUE; 7. CD5-3B-GOLD, GENESIS LG NARROW 3" PERF BOTTOM GOLD; 8. CD5-3B-GREEN, GENESIS LG NARROW 3" PERF BOTTOM GREEN; 9. CD5-3B-PURPLE, GENESIS LG NARROW 3" PERF BOTTOM PURPLE; 10. CD5-3B-RED, GENESIS LG NARROW 3" PERF BOTTOM RED; 11. CD5-3C-BLACK, GENESIS LG NARROW 3" SOLID BOTTOM BLACK; 12. CD5-3C-BLUE, GENESIS LG NARROW 3" SOLID BOTTOM BLUE; 13. CD5-3C-GOLD, GENESIS LG NARROW 3" SOLID BOTTOM GOLD; 14. CD5-3C-GREEN, GENESIS LG NARROW 3" SOLID BOTTOM GREEN; 15. CD5-3C-PURPLE, GENESIS LG NARROW 3" SOLID BOTTOM PURPLE; 16. CD5-3C-RED, GENESIS LG NARROW 3" SOLID BOTTOM RED; 17. CD5-61B-BLACK, GENESIS LG NARROW 6" PERF BOTTOM BLACK; 18. CD5-61B-BLUE, GENESIS LG NARROW 6" PERF BOTTOM BLUE; 19. CD5-61B-GOLD, GENESIS LG NARROW 6" PERF BOTTOM GOLD; 20. CD5-61B-GREEN, GENESIS LG NARROW 6" PERF BOTTOM GREEN; 21. CD5-61B-PURPLE, GENESIS LG NARROW 6" PERF BOTTOM PURPLE; 22. CD5-61B-RED, GENESIS LG NARROW 6" PERF BOTTOM RED; 23. CD5-61C-BLACK, GENESIS LG NARROW 6" SOLID BOTTOM BLACK 24. CD5-61C-BLUE, GENESIS LG NARROW 6" SOLID BOTTOM BLUE; 25. CD5-61C-GOLD, GENESIS LG NARROW 6" SOLID BOTTOM GOLD; 26. CD5-61C-GREEN, GENESIS LG NARROW 6" SOLID BOTTOM GREEN; 27. CD5-61C-PURPLE, GENESIS LG NARROW 6" SOLID BOTTOM PURPLE; 28. CD5-61C-RED, GENESIS LG NARROW 6" SOLID BOTTOM RED.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

All lot numbers are involved. UDI/DI: 1. CD5-3B - 10885403019548; 2. CD5-3C - 10885403019555; 3. CD5-61B - 10885403019579; 4. CD5-61C - 10885403019586; 5. CD5-3B-BLACK - 10885403197970; 6. CD5-3B-BLUE - 10885403197949; 7. CD5-3B-GOLD - 10885403197925; 8. CD5-3B-GREEN - 10885403197956; 9. CD5-3B-PURPLE - 10885403197932; 10. CD5-3B-RED - 10885403197963; 11. CD5-3C-BLACK - 10885403196515; 12. CD5-3C-BLUE - 10885403196485; 13. CD5-3C-GOLD - 10885403196461; 14. CD5-3C-GREEN - 10885403196492; 15. CD5-3C-PURPLE - 10885403196478; 16. CD5-3C-RED - 10885403198700; 17. CD5-61B-BLACK - 10885403198038; 18. CD5-61B-BLUE - 10885403198007; 19. CD5-61B-GOLD - 10885403197987; 20. CD5-61B-GREEN - 10885403198014; 21. CD5-61B-PURPLE - 10885403197994; 22. CD5-61B-RED - 10885403198021; 23. CD5-61C-BLACK - 10885403196577; 24. CD5-61C-BLUE - 10885403196546; 25. CD5-61C-GOLD - 10885403196522; 26. CD5-61C-GREEN - 10885403196553; 27. CD5-61C-PURPLE - 10885403196539; 28. CD5-61C-RED - 10885403196560.