CardioMEMS HF System Hospital Electronics System (HES)… recallZ-1221-2023 · St. Jude Medical · initiated Feb 7, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from St. Jude Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
FDA root cause category: Device Design
What the company is doing
Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to continue to follow CardioMEMS Instructions for Use (IFU). Abbott will notify patients of the emissions issue. If you suspect a CardioMEMS Electronics System has interfered with another medical device during use, report the event to Abbott Remote Care Technical Support at 1-844-MYCMEMS (692-6367) (US). Customers were also instructed to complete and return the acknowledgement and receipt form. Abbott notified patients on about 02/07/2023 via mailed letter. Customers were instructed that they could continue use of their CardioMEMS System and that their doctor has been notified of the issue as well. If there is suspicion that interference may be occurring during a sensor reading or any other quality issues, please report it to: Abbott Technical Support at 1-844-MYCMEMS (692-6367) (US); Monday-Friday: 8am-8pm EST
Recall details
- Recall number
- Z-1221-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 7, 2023
- Posted by the FDA
- Mar 3, 2023
- Recalling firm
- St. Jude Medical
- Quantity
- 1,905 units
- Product code
- MOM
- FDA submissions
- P100045
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Product description
CardioMEMS HF System Hospital Electronics System (HES), Model number CM3000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.
Code information (models, lots, serial numbers)
Show the full code information (23k characters)
UDI/DI 05415067029331, Serial Numbers: M140100179, M140100180, M140100181, M140100182, M140100183, M140100184, M140100185, M140100186, M140200187, M140200188, M140200190, M140200191, M140200193, M140200196, M140200204, M140200205, M140200213, M140200215, M140200216, M140200218, M140200220, M140200221, M140200223, M140200224, M140200226, M140200228, M140200230, M140200233, M140200234, M140200235, M140200237, M140200238, M140200240, M140200243, M140200245, M140400469, M140400470, M140400471, M140400472, M140400474, M140600522, M140600530, M140600532, M140600536, M140600537, M140600538, M140600541, M140600542, M140600543, M140600545, M140600546, M140700687, M140700688, M140700689, M140700690, M140700691, M140700692, M140700693, M140700694, M140700698, M140700699, M140700700, M140700701, M140700703, M140700705, M140700710, M140700712, M140700713, M140700714, M140700715, M140700717, M140700718, M140700721, M140700722, M140700725, M140700726, M140700799, M140700800, M140700801, M140700804, M140700806, M140700817, M140700818, M140800820, M140800821, M140800822, M140800825, M140800827, M140800828, M140800829, M140800833, M140800835, M140800837, M140800838, M140800839, M140800841, M140800842, M140800846, M140800848, M140800849, M140800850, M140800853, M140800855, M140800856, M140800859, M140800861, M140800862, M140800864, M140800867, M140800868, M140800869, M140800872, M140800874, M140800876, M140800877, M140800879, M140800880, M140800881, M140800883, M140800886, M140800887, M140800890, M140800891, M140800895, M140800896, M140800918, M140800919, M140800922, M140800925, M140800926, M140901157, M140901158, M140901159, M140901160, M140901161, M140901162, M140901163, M140901164, M140901166, M140901169, M140901172, M140901174, M140901175, M140901177, M140901184, M140901185, M140901186, M141001337, M141001339, M141001341, M141001344, M141001345, M141001353, M141001354, M141001355, M141001356, M141001358, M141001359, M141001360, M141001361, M141001362, M141001363, M141001366, M141201849, M141201857, M141201859, M141201861, M141201862, M141201863, M141201865, M141201866, M141201867, M141201869, M141201870, M141201871, M141201873, M141201874, M141201876, M141201879, M141201880, M141201882, M141201883, M141201885, M141201886, M141201887, M141201889, M141201890, M141201891, M141201893, M141201894, M141201895, M141201897, M141201898, M141201899, M141201900, M141201901, M141201902, M141201903, M141201904, M141201906, M150102237, M150102238, M150102239, M150102241, M150102243, M150102245, M150102246, M150102248, M150102253, M150102254, M150102255, M150102257, M150102259, M150102260, M150102261, M150102263, M150102265, M150102266, M150202497, M150202498, M150202499, M150202502, M150202503, M150202504, M150202505, M150202507, M150202509, M150202512, M150202514, M150202516, M150202526, M150302647, M150302649, M150302650, M150302652, M150302655, M150402907, M150402911, M150402912, M150402914, M150402916, M150402918, M150402919, M150402920, M150402921, M150402922, M150402924, M150402925, M150402926, M150402927, M150402929, M150402931, M150402932, M150402934, M150402935, M150402937, M150402938, M150402941, M150402942, M150402943, M150402944, M150402945, M150402946, M150403187, M150403188, M150403189, M150403190, M150403191, M150403192, M150403193, M150403194, M150503377, M150503378, M150503379, M150503380, M150503381, M150503384, M150503385, M150503390, M150503391, M150503393, M150503397, M150503398, M150503400, M150503401, M150503404, M150503407, M150503409, M150503410, M150503411, M150503413, M150503414, M150503415, M150503416, M150503419, M150503420, M150503423, M150503424, M150704378, M150704379, M150704380, M150704381, M150704383, M150704384, M150704385, M150704386, M150704387, M150704389, M150704391, M150704393, M150704397, M150704398, M150704399, M150704401, M150704403, M150704405, M150704407, M150704412, M150704413, M150704414, M150704416, M150704417, M150704418, M150704419, M150704420, M150704421, M150704422, M150704423, M150704424, M150704426, M150804897, M150804899, M150804901, M150804903, M150804904, M150804906, M150804907, M150804908, M150804909, M150804910, M150804911, M150804912, M150804914, M150804915, M150804916, M150804917, M150804918, M150804919, M150804920, M150804921, M150804922, M150804923, M150804924, M150804926, M150804927, M150804929, M150804932, M150804933, M150804934, M150804935, M150804936, M150804937, M150804938, M150804939, M150804943, M150804944, M150804945, M150804947, M150804948, M150804949, M150804950, M150804951, M150804952, M150804953, M150804954, M150804955, M150905408, M150905410, M150905411, M150905412, M150905413, M150905414, M150905415, M150905416, M150905417, M150905418, M150905419, M150905420, M150905424, M150905426, M150905427, M150905428, M150905429, M150905430, M150905431, M150905432, M150905433, M150905435, M150905717, M150905718, M150905719, M150905721, M150905722, M150905724, M150905728, M150905729, M150905730, M150905731, M150905732, M150905734, M150905739, M150905740, M150905741, M150905742, M150905745, M151005983, M151005988, M151005989, M151005994, M151005995, M151106357, M151106359, M151106362, M151106363, M151106364, M151106366, M151106367, M151106368, M151106369, M151106371, M151106372, M151106375, M151106376, M151106381, M151106382, M151106385, M151106387, M151106389, M151106390, M151106391, M151106392, M151106393, M151106394, M151106396, M151106397, M151106398, M151106399, M151106400, M151106401, M151106403, M151106404, M151106405, M151206527, M151206528, M151206530, M151206531, M151206533, M151206537, M151206539, M151206540, M151206541, M151206542, M151206544, M151206545, M151206546, M151206548, M151206549, M151206550, M151206551, M151206554, M151206556, M151206558, M151206559, M151206560, M151206561, M151206563, M151206566, M160307210, M160307211, M160307214, M160307219, M160307221, M160307223, M160307224, M160307226, M160307227, M160307234, M160707625, M160807677, M160807678, M160807679, M160807680, M160807682, M160807683, M160807684, M160807685, M160807687, M160807690, M160807692, M161107927, M161107928, M161107929, M161107930, M161107931, M161107932, M161107933, M161107934, M161107935, M161107936, M161107937, M161107938, M161107939, M161107940, M161107941, M161107942, M161107943, M161107944, M161107945, M161207947, M161207948, M161207953, M161207955, M161207956, M161207957, M161207958, M161207959, M161207960, M161207961, M161207962, M161207963, M161207964, M161207965, M161207968, M161207969, M161207970, M161207972, M161207973, M161207974, M161207975, M161207976, M161207978, M161207980, M161207981, M161207983, M161207984, M161207985, M161208049, M161208065, M161208066, M161208067, M161208068, M161208069, M161208070, M161208071, M161208072, M161208073, M161208074, M161208075, M161208076, M161208077, M161208078, M161208079, M161208080, M161208081, M161208082, M161208084, M170108086, M170108087, M170108088, M170108089, M170108090, M170108091, M170108092, M170108093, M170108095, M170108096, M170108097, M170108098, M170108100, M170108101, M170108102, M170108103, M170108104, M170108106, M170108107, M170108108, M170108109, M170108110, M170108111, M170108112, M170108113, M170108114, M170108115, M170108119, M170108120, M170108122, M170108123, M170108124, M170108125, M170108126, M170108127, M170108130, M170108131, M170108132, M170108133, M170108134, M170108135, M170108136, M170108137, M170108138, M170108140, M170108141, M170108142, M170108144, M170108145, M170108146, M170108147, M170108148, M170108149, M170108150, M170108153, M170108154, M170108155, M170108156, M170108157, M170208158, M170208159, M170208160, M170208161, M170208162, M170208163, M170208164, M170208165, M170208166, M170208167, M170208168, M170208169, M170208170, M170208171, M170208172, M170208173, M170208174, M170208175, M170208176, M170208177, M170208178, M170208179, M170208181, M170208182, M170208183, M170208184, M170208185, M170208187, M170308188, M170308190, M170308191, M170308192, M170308193, M170308194, M170308195, M170308196, M170308197, M170308200, M170308201, M170308202, M170308203, M170308204, M170308205, M170308206, M170308207, M170308208, M170308209, M170308210, M170308211, M170308212, M170308213, M170308214, M170308215, M170408216, M170408217, M170408218, M170408219, M170408220, M170408222, M170408223, M170408224, M170408225, M170408226, M170408227, M170408228, M170408229, M170408230, M170408231, M170408232, M170408233, M170408234, M170408235, M170408240, M170408241, M170408242, M170408243, M170408244, M170408246, M170408249, M170408250, M170408252, M170408253, M170408254, M170408255, M170408256, M170408257, M170408259, M170508260, M170508261, M170508275, M170608293, M170608294, M170608295, M170608296, M170608297, M170708308, M170708309, M170708310, M170708311, M170708313, M170708314, M170708315, M170708316, M170708317, M170708321, M170708323, M170708339, M170808325, M170808326, M170808328, M170808329, M170808336, M170808337, M170808338, M170808340, M170808341, M170808342, M170808344, M170808345, M170808346, M170808347, M170808352, M170808353, M170808357, M170908358, M170908360, M170908361, M170908362, M170908364, M170908366, M170908367, M170908368, M170908369, M170908370, M170908371, M170908372, M170908373, M170908374, M170908377, M170908378, M170908380, M170908383, M170908384, M170908387, M170908389, M170908390, M170908392, M170908393, M170908394, M170908395, M170908396, M171108397, M171108398, M171108399, M171108400, M171108402, M171108403, M171108406, M171108407, M171108408, M171108409, M171108410, M171108412, M171108413, M171108414, M171108415, M171108416, M171108417, M171108419, M171108422, M171108423, M171108424, M171108425, M171108426, M171108427, M171108428, M171108429, M171108430, M171108431, M171108432, M171108433, M171108436, M171108437, M171108438, M171108439, M171108440, M171108441, M171108442, M171108443, M180108444, M180108445, M180108451, M180108453, M180108455, M180108456, M180108457, M180108459, M180108460, M180108462, M180108464, M180108465, M180108467, M180108468, M180108469, M180108470, M180108471, M180108472, M180108474, M180208475, M180208476, M180208479, M180208480, M180208481, M180208482, M180208483, M180208484, M180208485, M180208488, M180208489, M180208491, M180208492, M180208493, M180208494, M180208495, M180208497, M180208498, M180208499, M180208501, M180208502, M180208503, M180208504, M180208505, M180208506, M180208507, M180208509, M180208510, M180208511, M180208512, M180208514, M180208517, M180308526, M180308527, M180308529, M180308535, M180308537, M180308539, M180408540, M180408543, M180408544, M180408545, M180408546, M180408549, M180408551, M180408552, M180408553, M180408554, M180408555, M180408556, M180408557, M180408562, M180408574, M180408575, M180408576, M180408577, M180408578, M180408579, M180408580, M180408583, M180408584, M180408585, M180408586, M180408587, M180408588, M180408589, M180408590, M180408591, M180508593, M180508596, M180508598, M180508600, M180508601, M180508602, M180508603, M180508604, M180508605, M180508606, M180508607, M180508608, M180508611, M180508617, M180508619, M180508620, M180508627, M180508629, M180508630, M180508631, M180608632, M180608634, M180608635, M180608637, M180608638, M180608639, M180608642, M180608643, M180608644, M180608645, M180608646, M180608647, M180608648, M180608649, M180608651, M180608652, M180608653, M180608655, M180608656, M180608657, M180608658, M180608659, M180608660, M180608661, M180608662, M180608663, M180608667, M180608668, M180608669, M180608670, M180608671, M180608672, M180608673, M180608674, M180608677, M180608678, M180608679, M180608680, M180608681, M180608683, M180608684, M180608689, M180608690, M180608691, M180608692, M180608693, M180608694, M180608697, M180608698, M180608699, M180608700, M180608701, M180608702, M180608707, M180608708, M180608710, M180608711, M180608712, M180608713, M180608714, M180608717, M180608718, M180608719, M180608720, M180608721, M180608722, M180608723, M180608724, M180608725, M180608726, M180608727, M180608728, M180608730, M180608731, M180608732, M180608734, M180608735, M180608736, M180608737, M180608738, M180608739, M180608740, M180608743, M180608745, M180608746, M180608747, M180608748, M180608749, M180608750, M180608751, M180608752, M180608753, M180608754, M180608755, M180608756, M180608757, M180608758, M180608759, M180608760, M180608762, M180608763, M180608764, M180608765, M180608766, M180608768, M180608769, M180608770, M180608772, M180608773, M180608774, M180608775, M180608776, M180608777, M180608778, M180608779, M180608782, M180608783, M180608784, M180608785, M180608786, M180608787, M180608788, M180608789, M180608790, M180608792, M180608793, M180608794, M180608795, M180608796, M180608797, M180608798, M180608800, M180608802, M180608803, M180608804, M180608805, M180608807, M180608808, M180608809, M180608810, M180608811, M180608812, M180608813, M180608815, M180708816, M180708819, M180708820, M180708821, M180708822, M180708827, M180808831, M180808833, M180808834, M180808836, M180808843, M180808851, M180808852, M180808853, M180808854, M181008860, M181008861, M181008862, M181008863, M181008865, M181008866, M181008867, M181008868, M181008872, M181008873, M181008874, M181008875, M181008876, M181008877, M181008878, M181008882, M181008883, M181008884, M181008885, M181008887, M181008888, M181008889, M181008890, M181008891, M181008892, M181008893, M181008894, M181008895, M181008896, M181008897, M181008898, M181008899, M181108904, M181108905, M181108907, M181108908, M181108910, M181108912, M181108913, M181208914, M181208916, M181208917, M181208918, M181208919, M181208920, M181208921, M181208922, M181208923, M181208924, M190108925, M190108927, M190108929, M190108930, M190108931, M190108933, M190108934, M190108935, M190108936, M190108956, M190208964, M190208965, M190208966, M190208967, M190208968, M190208969, M190208970, M190208971, M190208972, M190208973, M190208974, M190208975, M190208987, M190208988, M190208989, M190208990, M190208992, M190208993, M190208994, M190208995, M190208996, M190300005, M190300006, M190300007, M190300008, M190300009, M190300010, M190300011, M190300014, M190300015, M190300016, M190300017, M190300018, M190300019, M190400005, M190400006, M190400007, M190400008, M190400010, M190500019, M190500042, M190500043, M190500052, M190600005, M190700001, M190700002, M190700003, M190700004, M190700005, M190800008, M190800009, M190800011, M190800012, M190800013, M190800014, M190800015, M190800016, M190800017, M190800018, M190800027, M190800028, M190800029, M190800030, M190800031, M190900001, M190900002, M190900005, M190900024, M190900025, M190900026, M190900028, M190900032, M190900034, M190900035, M191000001, M191000003, M191000004, M191000005, M191000006, M191000007, M191000008, M191000009, M191000010, M191000011, M191100008, M191100009, M191100010, M191100011, M191200001, M191200002, M191200003, M191200004, M191200005, M200100001, M200100003, M200100004, M200100005, M200100006, M200100007, M200200001, M200200002, M200200003, M200200004, M200200009, M200300001, M200300006, M200300009, M200400001, M200400002, M200400004, M200400005, M200400017, M200400018, M200400019, M200400023, M200400024, M200500002, M200500003, M200500004, M200500005, M200500006, M200500007, M200500008, M200500009, M200500010, M200600001, M200600002, M200600003, M200600004, M200600005, M200600006, M200600007, M200600008, M200600011, M200600012, M200600013, M200600015, M200600017, M200600020, M200600021, M200600022, M200600023, M200700001, M200700007, M200700015, M200700026, M200700027, M200700028, M200700029, M200700030, M200700031, M200700032, M200800001, M200800002, M200800003, M200800004, M200800005, M200800006, M200800007, M200800008, M200800009, M200800010, M200800011, M200800012, M200800013, M200800014, M200800015, M200800016, M200800017, M200900001, M200900002, M200900003, M200900004, M200900005, M200900006, M200900007, M200900008, M200900009, M200900010, M200900011, M200900012, M200900013, M200900014, M200900015, M200900016, M200900017, M200900018, M200900019, M200900020, M200900021, M201000003, M201000004, M201000005, M201000006, M201000007, M201000008, M201000009, M201000010, M201000011, M201000012, M201000013, M201000015, M201000018, M201000019, M201000020, M201000021, M201100001, M201100002, M201100003, M201100004, M201100005, M201100006, M201100007, M201100008, M201100059, M201100060, M201100062, M201100063, M201100064, M201100065, M201100066, M201100067, M201100068, M201200001, M201200002, M201200003, M201200005, M201200006, M210100001, M210100002, M210100003, M210100005, M210100006, M210100007, M210100008, M210100010, M210100011, M210100012, M210100013, M210100014, M210100015, M210100016, M210100017, M210100019, M210100020, M210100021, M210200001, M210200002, M210200003, M210200004, M210200005, M210200006, M210200007, M210200008, M210200009, M210200010, M210200012, M210200013, M210200014, M210200015, M210200016, M210200017, M210200018, M210200019, M210200025, M210200030, M210200031, M210200032, M210200033, M210200034, M210200035, M210300003, M210300004, M210300005, M210300006, M210300007, M210300009, M210300010, M210300011, M210300012, M210300013, M210300015, M210300016, M210300018, M210300019, M210300020, M210300022, M210300023, M210300024, M210300025, M210400007, M210400008, M210400009, M210400010, M210400011, M210400012, M210400013, M210400014, M210400015, M210400027, M210400028, M210400029, M210400030, M210400031, M210400032, M210400033, M210400034, M210400035, M210400036, M210400037, M210400038, M210400039, M210400040, M210400041, M210400042, M210400043, M210400044, M210400045, M210400046, M210400047, M210400048, M210400049, M210400050, M210500001, M210500003, M210500004, M210500005, M210500006, M210500007, M210500008, M210500009, M210500010, M210500011, M210500012, M210500013, M210500014, M210500015, M210500016, M210500017, M210500018, M210500019, M210500020, M210500021, M210500022, M210500023, M210500024, M210500025, M210500026, M210500027, M210500028, M210500029, M210500030, M210500031, M210500032, M210500033, M210500034, M210500036, M210500037, M210500039, M210500040, M210500041, M210800002, M210800003, M210800005, M210800006, M210800007, M210800008, M210800009, M210800010, M210800011, M210800012, M210800013, M210800014, M210800015, M210800016, M210800017, M210800019, M210800020, M210800021, M210800022, M210800023, M210800024, M210800025, M210800026, M210800027, M210800028, M210800029, M210800030, M210800031, M210800032, M210800033, M210900001, M210900003, M210900004, M210900005, M210900006, M210900007, M210900008, M210900010, M210900011, M210900012, M210900013, M210900014, M210900015, M210900016, M210900017, M210900018, M210900019, M210900020, M210900021, M210900022, M210900023, M210900024, M210900025, M211000001, M211000003, M211000004, M211000005, M211000007, M211000008, M211000009, M211000010, M211000011, M211000012, M211000013, M211000014, M211200001, M211200002, M211200003, M211200004, M211200005, M140600524, M140700798, M140700803, M140800826, M140800866, M140901181, M140901183, M141201855, M141201864, M150202510, M150202513, M150402908, M150402910, M150402933, M150503406, M150503408, M150503425, M150905423, M151005977, M151005980, M151005981, M151005982, M151005984, M151005985, M151005986, M151005987, M151005991, M151005992, M151005993, M151206552, M151206557, M160307208, M160307222, M160307225, M160307231, M160307235, M160707618, M160707619, M160707621, M160707622, M160707631, M160707632, M160707633, M160707636, M161208051, M170108105, M170608276, M170608278, M170608279, M170608280, M170608281, M170608283, M170608284, M170608288, M170608291, M170608292, M170608299, M170608300, M170608304, M170608306, M170608307, M170708322, M170808330, M170808331, M170808332, M170808333, M170808334, M170808335, M170808348, M170808349, M170808350, M170808351, M171108405, M171108435, M180208513, M180308522, M180308524, M180308528, M180308530, M180308531, M180308532, M180308533, M180308534, M180408558, M180408559, M180408560, M180408561, M180408592, M180508594, M180508595, M180508613, M180508614, M180508615, M180508622, M180508623, M180508624, M180608640, M180608664, M180608665, M180608666, M180608675, M180608676, M180608685, M180608686, M180608704, M180608705, M180608706, M180608715, M180608716, M180708824, M180708825, M180708826, M180808828, M180808829, M180808838, M180808839, M180808842, M180808844, M180808846, M180808847, M180808848, M180808849, M180808850, M180908856, M180908858, M180908859, M181008869, M181008879, M181008880, M181008901, M181008902, M181008903, M190108937, M190108938, M190108939, M190108940, M190108941, M190108942, M190108943, M190108945, M190108947, M190108948, M190108949, M190108950, M190108951, M190108952, M190108953, M190108954, M190300001, M190300003, M190300004, M190400001, M190400003, M190400004, M190500001, M190500002, M190500003, M190500004, M190500005, M190500006, M190500007, M190500008, M190500009, M190500016, M190500017, M190500018, M190500021, M190500022, M190500044, M190500045, M190500046, M190500047, M190500048, M190500049, M190500050, M190500051, M190600001, M190600002, M190600006, M190600007, M190600008, M190600009, M190600010, M190600011, M190600012, M190600013, M190600014, M190600015, M190600017, M190600018, M190600019, M190600020, M190600021, M190600023, M190600026, M190600027, M190600028, M190600029, M190700006, M190700007, M190700008, M190700009, M190700010, M190700011, M190700012, M190700013, M190700014, M190700015, M190700016, M190700017, M190700018, M190700019, M190700020, M190700021, M190700022, M190700023, M190700024, M190700025, M190700026, M190700027, M190700029, M190700030, M190700031, M190700032, M190700033, M190700034, M190700035, M190700036, M190800002, M190800003, M190800004, M190800005, M190800006, M190800007, M190800019, M190800020, M190800021, M190800022, M190800023, M190800024, M190800025, M190800026, M190900039, M190900041, M200700002, M200700003, M200700004, M200700005, M200700006, M200700008, M200700009, M200700010, M200700011, M200700012, M200700013, M200700014, M200700016, M200700017, M200700018, M200700019, M200700020, M200700021, M200700022, M200700024, M200700025, M210200020, M210200021, M210200022, M210200023, M210200024, M210200026, M210200027, M210200028, M210200029, M210200038, M210200039, M210300001, M210300002, M210300014, M210300017, M210400001, M210400002, M210400003, M210400004, M210400006, M210400018, M210400019, M210400020, M210400021, M210400022, M210400023, M210400025, M210400026, M210500042, M210500043, M210600001, M210600002, M210600003, M210600004, M210600005, M210600006, M210600007, M210800035, M210800036, M210800037, M210800038, M210800039, M210800040, M210800041, M210800042, M210800043, M210800044, M210800045, M210800046, M211000023, M211000024, M211000025, M220600002, M220600003, M220600004, M220600005, M220600006, M220600008, M220600011, M220600034, M220600036, M220600037, M220600038, M220600039, M220600040, M220600041, M220600042, M220700189, M220700192, M161208048
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CardioMEMS™ CM3000Cardiomems, Inc. · System, hemodynamic, implantable3 recalls65 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91610).
- Z-1218-2023 · Open, ClassifiedCardioMEMS HF System Patient Electronics System (PES), Model number CM1000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and man…
- Z-1219-2023 · Open, ClassifiedCardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and man…
- Z-1220-2023 · Open, ClassifiedCardioMEMS HF System Patient Electronics System (PES), Model number CM1100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and man…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗