Skip to content

Baxter MiniCap Disconnect Cap with Povidone-Iodine… recallZ-1213-2023 · Baxter Healthcare Corporation · initiated Feb 1, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.

FDA root cause category: Process control

What the company is doing

Baxter Healthcare initiated the recall by issuing an Urgent Medical Device Recall notice to its consignees by letter on 02/01/2023. The notice requested the following from customers: 1. Check your stock and set aside all product with the affected lot numbers. 2. Contact Baxter HomeCare Services to arrange for return of impacted product and for ordering replacement product. 3. If all of your product is impacted by this recall and you are unable to get unaffected lot numbers, resulting in a significant delay in therapy, please follow the below steps: a. Contact your clinic for clinical guidance and check each pouch and do not use the MiniCap contained inside of any pouches that appear to have damaged seals. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday." Recall was expanded and Baxter Healthcare sent an updated notification to customers and patients on 05/09/2023. The letter instructed consignees and patients to immediately cease use of the affected product, arrange for the return of affected product and assess each patient's individual next steps based on clinical evaluation. Customers were also instructed to complete the customer response form via the portal (for products directly received from Baxter) or through the distributor. If the products were further distributed, please notify customers.

Recall details

Recall number
Z-1213-2023
Status
Open, Classified
Initiated by the company
Feb 1, 2023
Posted by the FDA
Mar 3, 2023
Recalling firm
Baxter Healthcare Corporation
Quantity
33,866,400 units
Product code
KDJ
FDA submissions
K152129
Distribution
US Nationwide distribution and Canada.

Product description

Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administration set for peritoneal dialysis

Code information (models, lots, serial numbers)

UDI/DI 00085412007694, Lot numbers: GD912051, exp 31-Mar-2024; GD912068, exp 30-Apr-2024; GD912099, exp 30-Apr-2024; GD912112, exp 30-Apr-2024; GD912143, exp 30-Apr-2024; GD912204, exp 31-May-2024. Expansion on 05/10/2023: Product Code 5C4466P with all lots numbers manufactured prior to 09/30/2022.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K152129 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page