Cardiosave Rescue. Model Number: 0998-00-0800-83 recallZ-1232-2023 · Datascope Corp. · initiated Feb 7, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
FDA root cause category: Equipment maintenance
What the company is doing
Firm notified customers on February 7, 2023, in an Urgent Medical Device Correction letter which discussed four issues in total - this issue was Issue #4. Customers were instructed to contact their Datascope/Getinge service representative if they observe any visual damage to the O-ring installed as part of the quick disconnect fitting. Datascope/Getinge is currently updating the annual Preventive Maintenance instruction to include replacement of the quick disconnect fitting O-ring. If you have any questions, please contact your Datascope/Getinge representative or call the Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). On 11/04/2025 the recall was expanded, and consignees were notified via letter. The notification was for awareness to the update of Datascope s preventative maintenance in the IFU addendum, which was provided to consignees. The IFU Addendum contains an updated Schedule B Preventive Maintenance Schedule which calls for the Quick Disconnect Fitting O-Ring to be replaced and to lubricate the console release latch every 12 months or 2500 hours. The Service Manual has contained this update to Preventive Maintenance for the Helium O-Ring since June 2023. Consignees were instructed to review the IFU Addendum and local Preventive Maintenance Schedule and update their Preventive Maintenance schedule accordingly.
Recall details
- Recall number
- Z-1232-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 7, 2023
- Posted by the FDA
- Mar 6, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 47 units (44 US, 3 OUS)
- Product code
- DSP
- FDA submissions
- K181122
- Distribution
- Nationwide US distribution. Worldwide international distribution.
Product description
Cardiosave Rescue. Model Number: 0998-00-0800-83.
Code information (models, lots, serial numbers)
Model Number: 0998-00-0800-83. UDI-DI: 10607567108407. All serial numbers.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91642).
- Z-1231-2023 · Open, ClassifiedCardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page