Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model… recallZ-1229-2023 · Datascope Corp. · initiated Feb 7, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.
FDA root cause category: Under Investigation by firm
What the company is doing
URGENT MEDICAL DEVICE CORRECTION dated 2/7/23 were sent to customers. There have been reported failures of the high pressure helium regulator which may cause a helium leak in the Cardiosave Hospital Cart. The high pressure helium regulator is located in the Cardiosave Hospital Cart and regulates the helium pressure of the external helium supply. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir. Please refer to images under Issue 3 for reference. Issue 3: Leak in the High Pressure Helium Regulator Identification of the issue: There have been reported failures of the high pressure helium regulator which may cause a helium leak in the Cardiosave Hospital Cart. The high pressure helium regulator is located in the Cardiosave Hospital Cart and regulates the helium pressure of the external helium supply. In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir. Please refer to images below for reference. Datascope/Getinge has received 51 reported complaints of helium leak as a result of high pressure helium regulator failure over a 2 year period. There have been 0 adverse events reported. Risk to Health: Should a Cardiosave s helium supply be depleted due to an impaired helium pressure regulator, therapy may be interrupted. As with any therapy interruption, the degree of subsequent hemodynamic stability is related to the patient s overall clinical condition, those critically ill are more vulnerable to clinical decline. The risk of therapy interruption from a depleted helium supply is mitigated by the advanced notice provided to the User (a minimum of approximately 24 hours). The Pump Console s internal helium
Recall details
- Recall number
- Z-1229-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 7, 2023
- Posted by the FDA
- Mar 6, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 8909 units
- Product code
- DSP
- FDA submissions
- K112372
- Distribution
- US Nationwide. Global Distribution.
Product description
Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, and 0998-00-0800-55
Code information (models, lots, serial numbers)
All Serial Numbers. Model Number: 0998-00-0800-31, UDI-DI:10607567109053; Model Number: 0998-00-0800-32, UDI-DI: 10607567111117; Model Number: 0998-00-0800-33, UDI-DI: 10607567109008; Model Number: 0998-00-0800-34, UDI-DI: 10607567111940; Model Number: 0998-00-0800-35, UDI-DI: 10607567109107; Model Number: 0998-00-0800-45, UDI-DI: 10607567108421; Model Number: 0998-00-0800-52, UDI-DI: 10607567108438; Model Number: 0998-00-0800-53, UDI-DI: 10607567108391; Model Number: 0998-00-0800-55, UDI-DI: 10607567108414; Model Number: 0998-00-0800-65, UDI-DI: 10607567113432
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
- CARDIOSAVE HYBRID TYPE I (AU) 0998-00-0800-31Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls232 reports
- CARDIOSAVE HYBRID TYPE D PLUG 0998-00-0800-33Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths230 reports
- CARDIOSAVE HYBRID TYPE J PLUG 0998-00-0800-32Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls2 deaths48 reports
- CARDIOSAVE HYBRID TYPE M PLUG 0998-00-0800-35Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls13 reports
- CARDIOSAVE HYBRID TYPE K PLUG 0998-00-0800-34Datascope Corp. · System, Balloon, Intra-Aortic And Control15 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91645).
- Z-1230-2023 · Open, ClassifiedCardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, and 0998-00-0800-85
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page