Luminos dRF Max with software VE10, VF10, VF11- A… recallZ-1167-2023 · Siemens Medical Solutions USA, Inc · initiated Dec 21, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential risk of collision with the ceiling, wall, or objects which may result in serious injury to staff or patients due to room configuration parameters of the system being set to default values
FDA root cause category: Software design
What the company is doing
***UPDATE*** Siemens issued Urgent Medical Device Correction letter (expansion) via CSAN XP056/22/S to affected customers on 2/24/2023. The CSAN alerts the users to the issue warning them to pay special attention to the travel range of system parts. The CSAN urges users to stop system movement to prevent collision by releasing the movement control or by activating the emergency stop button as described in the system Operator Manual. Field corrective action via XP057/22/S will initiate an onsite visit to the customer facilities by a Siemens Customer Service Engineer to check the room configuration parameters current set on the systems and, if necessary, restore them to the correct values. This service will be provided to the customers free of charge. Customers with software version VF11 will be addressed under a separate and to be determined UI (update to VF11H). If you have any technical questions or desire an earlier appointment, please contact our service organization at 1-800-888-7436. _______________________________ ___________________________________________ Siemens Medical Solutions issued Customer Safety Advisory Notification( CSAN XP053/22/S) beginning on December 21, 2022. Letter states reason for recall, health risk and action to take: When performing unit movements, the user should pay special attention to the travel range of the system parts and stop movement in time to prevent a collision with the ceiling, wall, or objects. System movement can be stopped by releasing the movement control or by activating the emergency stop button as described in the system Operator Manual. Customer Safety Advisory Notice (XP053/22/S) will be distributed to all affected customers. Field corrective action via XP054/22/S will initiate an onsite visit of a Siemens Customer Service Engineer to check the room configuration parameters currently set for your system and if necessary, restore them to the correct values. These visits will be free of charge. The Siemens Service Or
Recall details
- Recall number
- Z-1167-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 21, 2022
- Posted by the FDA
- Feb 23, 2023
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 2 units; Expanded recall: 53 units
- Product code
- OWB
- FDA submissions
- K173639
- Distribution
- US Nationwide distribution.
Product description
Luminos dRF Max with software VE10, VF10, VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model Number:10762471
Code information (models, lots, serial numbers)
UDI: 4056869009162 S/N: 5037 5951 Expanded Recall 2/24/23: 5069 5085 5209 5234 5340 5364 5634 5664 5682 5919 5920 5921 5947 5953 5954 5956 5963 7075 7098 7147 7162 7168 7172 7208 7266 7310 7319 7391 7404 7442 7460 7488 7545 7743 7771 8074 8093 8094 8183 8198 8223 8242 8244 8275 8320 8322 8334 8361 8371 8378 8385 8389
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Luminos Agile Max 10762472Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified3 recalls37 reports
Other devices under the same FDA submission
2 devices cleared under K173639 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LUMINOS Lotus Max 11574100Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system2 recalls7 reports
- Multitom Rax 10860700Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91649).
- Z-1165-2023 · Open, ClassifiedLuminos Agile Max with software VE10 & VF11- A diagnostic imaging system for radiographic and fluoroscopic studies Model Number: 10762472
- Z-1168-2023 · Open, ClassifiedUroskop Omnia Max with software VE10 & VF11- A diagnostic imaging system for radiographic and fluoroscopic studie Model Number:10762473
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page