Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model… recallZ-1277-2023 · Datascope Corp. · initiated Feb 28, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION dated 2/28/23 was sent to customers. User Actions to be taken: A review of our records indicate that you may have a Cardiosave IABP at your facility with an affected Safety Disk or affected spare Safety Disks or PIM assembly installed recently into Cardiosave IABP. 1. Please examine your inventory immediately to determine if you have any Cardiosave IABPs with the affected Safety Disks and/or PIM, or any affected Safety Disk and/or PIM assembly spare parts, per attachment 1. 2. You may continue to use any affected IABPs until replacement parts are installed. Please keep in mind the potential health risks cited above, particularly with prolonged heart rates above 100 beats per minute for three consecutive hours. When transporting patients outside the hospital environment, consider using an unaffected console if available. 3. Should you have any unused affected spare parts of the Safety Disk or PIM Assembly in your inventory, please remove them from areas of use and contact your local Datascope/Getinge Customer Service department at 888-9GETUSA (888 943-8872) to request a return authorization (RMA) and shipping instruction to return the affected product for credit. 4. If you installed an affected spare part into one of your IABPs the Safety Disk will need to be replaced. 5. If you purchased spare parts and they are no longer in your inventory and you are unsure of their location, please contact Getinge Service. Getinge Service will schedule an on-site visit to inspect your IABP/(s) at no cost to your facility. 6. Should you have any affected Cardiosave IABPs you will be contacted by a Getinge Service Representative to schedule an on-site visit to replace the Safety Disk, free of charge. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Technical Support at (888) 943-8872 (select option 4,2,1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone

Recall details

Recall number
Z-1277-2023
Status
Open, Classified
Initiated by the company
Feb 28, 2023
Posted by the FDA
Mar 22, 2023
Recalling firm
Datascope Corp.
Quantity
391 total IABPs
Product code
DSP
FDA submissions
K112372
Distribution
Distribution list not yet provided.

Product description

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Code information (models, lots, serial numbers)

Model Number: 0998-00-0800-75, UDI-DI: 10607567112312; Model Number: 0998-00-0800-83, UDI-DI: 10607567108407; Model Number: 0998-00-0800-85, UDI-DI: 10607567113449; Serial Numbers: CA237026J3 CH231276C6

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 91652).