Skip to content

Atellica CH Triglycerides (concentrated)-In vitro… recallZ-1270-2023 · Siemens Healthcare Diagnostics, Inc. · initiated Jan 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reagent Carryover from LDL Cholesterol Direct (DLDL), Total Protein II (TP), Triglycerides (Trig), and Triglycerides_2 (Trig_2) Impacting Magnesium (Mg) Results

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens Issued Urgent Field Notice to US customers ACHC23-01.B.US via FedEx beginning on 01/30/2023. In accordance with outside US country regulations, Siemens in-country personnel for the EU and non-EU regions follow local requirements for contacting their respective National Competent Authorities customers for distribution. and regional reporting to their local competent authorities beginning on 01/27/2023. The UMDC/UFSN provides instructions to all Atellica CH 930 customers to either separate the assays causing carryover by analyzer, if they have multiple analyzers, or by batch testing of Mg. In addition, the UMDC and UFSN indicates that SW v1.27 and higher will correct this issue. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director. Perform the instructions provided in the Additional Information section. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Recall details

Recall number
Z-1270-2023
Status
Open, Classified
Initiated by the company
Jan 27, 2023
Posted by the FDA
Aug 29, 2023
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
11824 units
Product code
CDT
FDA submissions
K133067
Distribution
Nationwide Foreign: Country Albania Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile China Colombia Croatia Cura¿ao Czech Republic Denmark Ecuador Egypt Finland France Germany Greece Guadeloupe Hong Kong Hungary India Ireland Israel Italy Japan Kuwait Libya Malaysia Mexico Netherlands New Zealand Norway Oman Pakistan Paraguay Peru Philippines Poland Portugal Qatar Republic Korea Romania Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan, Province of China Thailand Ukraine United Arab Emirates United Kingdom Great Britain Uruguay Vatikancity Vietnam

Product description

Atellica CH Triglycerides (concentrated)-In vitro diagnostic use in the quantitative measurement of triglycerides in human serum and plasma (lithium heparin, potassium EDTA) using the Atellica CH Analyzer SMN: 11097591

Code information (models, lots, serial numbers)

UDI: 00630414596495 All Lots

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 91665).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page