Skip to content

Fusion Bioline Vascular Graft, Part number M00201503046B0 recallZ-1245-2023 · Maquet Cardiovascular, LLC · initiated Jan 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.

FDA root cause category: Process control

What the company is doing

On January 30, 2023, the firm notified customers via Urgent Medical Device Removal letters. Customers are asked to examine your inventory immediately for lot 25162546. Please note, the Lot number appears on the device shelf boxes only. If you have the affected lot, please stop using and remove the device from areas of use. You are entitled to a replacement at no cost to your facility. You will receive replacement upon your acknowledgement that you have product for return. Distributors who have shipped this product should notify their customers. Please contact Getinge Customer Service at 888-943-8872 between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product.

Recall details

Recall number
Z-1245-2023
Status
Open, Classified
Initiated by the company
Jan 30, 2023
Posted by the FDA
Mar 10, 2023
Recalling firm
Maquet Cardiovascular, LLC
Quantity
1 (US)
Product code
DSY
FDA submissions
K131778
Distribution
US Nationwide distribution in the states of FL, IN, LA, MI, MO, NC, NV, NY, PA, TN, TX, VA, WA.

Product description

Fusion Bioline Vascular Graft, Part number M00201503046B0

Code information (models, lots, serial numbers)

UDI-DI 00384409005942 Lot 25162546

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

4 devices cleared under K131778 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates