Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro… recallZ-1258-2023 · Siemens Healthcare Diagnostics, Inc. · initiated Jan 30, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of erroneous patient results are major, since in a worst case scenario, a patient s tacrolimus dosing could be altered, resulting in potential organ failure and rejection (due to too low or no tacrolimus being administered) or toxicity (due to too much tacrolimus being administered). Siemens internal investigation of the Dimension TAC assay showed the worst case imprecision with patient samples at the low end of the Analytical Measurement Range (AMR). A patient sample at 2.0 ng/mL (2.6 nmol/L) recovered as 0.0 ng/mL (0.0 nmol/L) (100% negative bias), another patient sample at 2.1 ng/mL (2.7 nmol/L) recovered at 4.3 ng/mL (5.6 nmol/L) (103% positive bias).
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued to US Customers Urgent Medical Device Correction (DC-23-03.A.US) via FedEx beginning on 01/30/2023. OUS Customers sent Urgent Field Safety Notice (DC-23-03.A.OUS). Letter states reason for recall, health risk and action to take: Discontinue use and discard the lots listed in Table 1 and complete the attached Product Replacement form for no-charge product replacement. Please note, lot GA2286 is expired. Review your inventory to determine possible replacement needs and provide information to Siemens for reporting to the Authorities. Please review this letter with your Medical Director and the decision on a review of previously generated results. If you receive any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Remote Services Center or your local Siemens Technical Support Representative. Complete and return the Field Correction Effectiveness Check/Product Replacement Form attached to this letter within 30 days to Siemens Healthineers for reporting to the Authorities. Questions, contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers Technical Support Representative.
Recall details
- Recall number
- Z-1258-2023
- Status
- Open, Classified
- Initiated by the company
- Jan 30, 2023
- Posted by the FDA
- Mar 15, 2023
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 6035 units
- Product code
- MLM
- FDA submissions
- K150168
- Distribution
- Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile Colombia Country India Japan New Zealand Nicaragua Saudi¿Arabia¿¿¿ South Africa South Korea Spain¿¿¿¿¿¿¿¿¿¿ Sri Lanka Taiwan
Product description
Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus in human whole blood on the Dimension clinical chemistry system SMN: 10700795 (DF207)
Code information (models, lots, serial numbers)
UDI: 00842768035425GA228622101310700795840, 00842768035425GA304723021610700795840, 00842768035425GA317123062010700795840. Lot Numbers: GA2286, GA3047, GA3171
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Dimension® Flex® reagent cartridge TAC 10700795Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, TRACROLIMUS2 recalls121 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page