Sit to Stand STS500 Patient Lift recallZ-1205-2023 · Med-Mizer, Inc. · initiated Feb 3, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Med-Mizer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.
FDA root cause category: Device Design
What the company is doing
Med-Mizer notified distributors of this recall event by distributing an Urgent Medical Device Recall Notice on 02/03/2022 via Certified Mail. The MR600 and STS500 Patient Lifts are being subject to recall due to the potential for the nut securing the boom to become loose over time, which could result in the boom moving unexpectedly side to side or dropping the patient during operation. Customers are asked to inspect all affected devices at their facility to determine if the nut securing the boom is loose. If it is, immediately remove the device from service. Med-Mizer is providing customers with affected devices replacement parts in order to complete the correction. Customers are to complete the provided Correction Response Form to Med-Mizer and return via fax at (812)932-5678, by email at [email protected], or by mail at 80 Commerce Drive, Batesville, IN 47006. If your firm is a distributor, please provide this notice to your customers. Any questions can be directed to Med-Mizer by phone at (877)867-7365.
Recall details
- Recall number
- Z-1205-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 3, 2023
- Posted by the FDA
- Mar 2, 2023
- Recalling firm
- Med-Mizer, Inc.
- Quantity
- 37 units
- Product code
- FSA
- Distribution
- Awaiting consignee list from the recalling firm.
Product description
Sit to Stand STS500 Patient Lift
Code information (models, lots, serial numbers)
UDI-DI: 00852195007353; Serial No.: STS500000, STS500001, STS500002, STS500003, STS500004, STS500005, STS500006, STS500007, STS500008, STS500009, STS500010, STS500011, STS500012, STS500013, STS500014, STS500015, STS500016, STS500017, STS500018, STS500019, STS500020, STS500021, STS500022, STS500023, STS500024, STS500025, STS500026, STS500027, STS500028, STS500029, STS500030, STS500031, STS500032, STS500033, STS500034, STS500035, and STS500036.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Med-Mizer, Inc. MED-RISER STS500Med-Mizer, Inc. · Lift, Patient, Non-Ac-Powered1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91695).
- Z-1204-2023 · Open, ClassifiedMed-Riser MR600 Patient Lift
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.